I-SENS AutoChek t Owner's Booklet - Note:
www.i-sens.com 27
Comparing the Control Solution Test Results
Repeat the control test if the test result falls outside the range
printed on label of the test strip vial. Out of range results
may occur due to the following factors:
•
When the control solution vial was not shaken well,
•
When the control solution is past its expiration date or is
contaminated,
•
When the meter, the strip or the control solution were exposed to high or low
temperatures,
•
When the first few drops of the control solution were not discarded or the tip of
the vial was not wiped clean,
•
When the test strip is past its expiration date,
•
When the meter is not functioning properly.
Note:
The AutoChek t Control Solution can be purchased separately. Please contact
your authorized i-SENS sales representative.