TA01i is class B equipment and needs special precautions regarding EMC and needs to be installed and put into service according to EMC information
provided in user and maintenance manual.
Guidance and manufacturer’s declaration-Electromagnetic emissions
TA01i is intended for use in the electromagnetic environment specified below. The user of the TA01i should assure that it is used in such an
environment.
RF emissions CISPR 11
Group 1
Icare TA01i is battery operated and use RF
energy only for its internal function.
Therefore, its RF emissions are low and are
not likely to cause any interference in nearby
equipment.
RF emissions CISPR 11
Class B
Icare TA01i is suitable for use in all
establishments, including domestic
establishments and those directly connected
to public low-voltage power supply network
that supplies buildings used for domestic
purposes
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Voltage fluctuations flickering emissions
IEC 61000-3-3
NOT APPLICABLE
Guidance and manufacturer’s declaration-Electromagnetic immunity
Icare TA01i is intended for use in the electromagnetic enviroment specified below. The customers or users of Icare TA01i should assurre
that it is used in such enviroment
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment-
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
Transients/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
NOT APPLICABLE
Surge
IEC 61000-4-5
±1 kV for line(s) to line(s)
±2 kV for line(s) to earth
NOT APPLICABLE
Voltage dips, short interruption
and voltage variations on power
supply lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
NOT APPLICABLE
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.