PERFORMANCE DATA
The performance data is obtained from a clinical study, performed according to American National Standard ANSI Z80.10-2003 and International Standard
ISO 8612.2 for tonometers. The study was performed ain the Department of Ophthalmology, Helsinki University Central Hospital. In the study, 158
patients were measured. The mean paired difference and standard deviation (Goldmann-Icare) were -0.4 mmHg and 3.4 mmHg. A scattergram and Bland-
Altman plot of the results is shown below.
ACCESSORIES
SKU
PRODUCT DESCRIPTION
WEIGHT
DIMENSIONS
104
BOX OF 100 PROBES
89,10 g
5,3 x 10,3 x 3,6 cm
102
BOX OF 100 PROBES (ONLY USA)
89,10 g
5,3 x 10,3 x 3,6 cm
7210
PROBE BASE COLLAR
1g
11 mm x 10 mm
540
PROBE BASE
4g
7 mm x 38 mm
550
TABLE STAND
52g
73 mm x 52 mm x 38 mm
560
WRIST STRAP
3g
270 mm x 10 mm x 10 mm
500
ALUMINIUM CASE
700g
240 mm x 280 mm x 72 mm
7171
BATTERY COVER & SCREW
4g
42 mm x 17 mm x 13 mm
7000
THREAD PLATE
1g
4 mm x 10 mm x 3 mm
543
PROBE BASE CLEANING CONTAINER
3g
5,6 cm x 2 cm
561
562
563
564
SILICON GRIP - WHITE
SILICON GRIP - PINK
SILICON GRIP - GREEN
SILICON GRIP - BLUE
26g
135 mm x 46 mm x 23 mm
MAINTENANCE
Follow local regulations and recycling instructions regarding the disposal or recycling of the Icare tonometerand accessories.
WARNING
The tonometer should only be opened by qualified service personnel. It contains no userserviceable parts, apart from the batteries and a probe base. The
Icare tonometerrequires no routine servicing or calibration other than changing the batteries at least every 12 months or changing or cleaning the probe
base. If servicing is necessary, contact qualified service personnel or your local Icare representative.