Battery compartment cover
Replacing the probe base
Replace the probe base every twelve months. Clean or replace the probe base if the error messages E01 or E03 are displayed.
Instructions for replacing the probe base:
• Turn off the tonometer.
• Unscrew the probe base collar and put it in a safe place.
• Remove the probe base by tilting the tonometer downwards and use your fingers to pull the probe base
out of the tonometer.
• Insert a new probe base into the tonometer.
• Screw the collar in, to lock the probe base.
Cleaning the probe base
You can reuse the probe base after careful cleaning. Clean the probe base every six months. Clean or
replace the probe base if the error messages E01 or E03 are displayed.
Instructions for cleaning the probe base:
• Fill the probe base cleaning container or other clean container with 100% isopropyl alcohol.
• Turn the power off.
• Unscrew the probe base collar.
• Invert the probe base over the container, drop in the probe base into the container and let soak for 5-30 minutes.
• Remove the probe base from alcohol.
• Dry the probe base by blowing clean canned or compressed air into the hole in the probe base. This will additionally remove possible residual dirt.
• Insert the probe base into the tonometer.
• Screw the collar in, to lock the probe base.
Cleaning the tonometer
WARNING
Never spray, pour or spill liquid onto the Icare tonometer, its accessories, connectors, switches or openings in the chassis. Dry any liquid on the surface of
the tonometer immediately.
Icare TA01’s surfaces have been tested and found chemically resistant to the following liquids:
• 100 % 2-propanol
• Mild soap solution
• 95% Pursept solution
Cleaning instructions for surfaces:
• Turn the power off.
• Dampen a soft cloth with one of the liquids mentioned above.
• Lightly wipe the surfaces of the tonometer with the soft cloth.
• Dry the surfaces with a dry soft cloth.
Replacing the batteries
Unscrew the battery compartment locking screw with a screwdriver or a small coin.
Remove the battery compartment cover. Remove the old batteries.
Insert a new set of batteries (four AA batteries). Do not use rechargeable batteries, since they may not function properly (the inner resistance of some
rechargeable batteries is too high). Insert the batteries in accordance with the diagrams inside the battery compartment, with the +terminals pointing
downwards on the display side of the tonometer (the rear side), and the -terminals pointing downwards on the measurement side (the front side).
Replace the battery compartment cover and secure it by screwing it in lightly using the coin or screwdriver. Take care not to use excessive force when
screwing the cover into place.