Interface 9834 Instruction Manual

Manual is about: High Level Output Indicator with Enhanced Performance Features

Summary of 9834

  • Page 1

    Instruction manual high level output indicator with enhanced performance features mode tare cal cal tare sp1 sp2 sp1 sp2 model 9834 disp reset rcal pk val 7401 e. Butherus dr. • scottsdale, az 85260 tel: (480) 948-5555 fax: (480) 948-1924 www.Interfaceforce.Com advanced force measurement.

  • Page 2

    Warranty and repair policy 1. Interface warrants that its instruments shall be free from defects in material and workmanship for one year under normal and proper use when correctly installed. 2. Any interface product which proves defective in material or in workmanship within one year from date of s...

  • Page 3

    Table of contents description and specifications ......................... 1 installation and wiring ........................................ 2 switch and function definitions....................... 6 operating instructions..................................... 10 calibration instructions ..............

  • Page 4

    Description the model 9834 is an enhanced performance, microprocessor-based digital indicator capable of interfacing to high-level voltage ( + 5 or + 10vdc) or current (4-20ma) output transducers. A 16-bit analog-to-digital converter accurately digitizes the input signal. A 12vdc@90ma, short circuit...

  • Page 5

    Installation and wiring installation the model 9834 enclosure is designated for panel mounting in a 1/8 din cutout. The cutout dimensions are shown below. To panel mount the 9834, perform the following steps. 1. Rotate the four pawl screws (outside screws in each corner) several turns counter-clockw...

  • Page 6

    Status leds (see status definition) case screw (1 of 4 inside screws) these 4 screws hold the front panel to the instrument case. Input / output connector (15-pin male d-sub) transducer connector (9-pin female d-sub) ac power connector (3-prong grounded iec connector) fuse holder (250v, 250ma, 5mm f...

  • Page 8

    Page 5 10v analog output adjust 20ma analog output adjust 4ma analog output adjust transformer figure 5. Location of switches, jumpers and potentiometers (top view) figure 6. Location of jumpers for ac line voltage selection (bottom view) 115 220 solder jumpers shown (2) for 115vac operation. For 23...

  • Page 9

    Page 6 switch definitions mode selection switch (s1) s1 is used to cycle through the various programmable functions of the 9834. The sequence is as follows. The designators, in ( ), identify text that will be displayed. The settable values are shown in { } factory default settings full scale cal swi...

  • Page 10

    Switch definitions (cont) increment or cal switch (s3) this switch serves two functions, depending on the present status of s1. If the instrument is in the mode selection sequence, this switch will increment the flashing digit. If the instrument is in its operating mode, s3 becomes an auto cal switc...

  • Page 11

    Function definitions filter value (filter) {00 to 99} the 9830 uses a filter based on exponential smoothing, a digital averaging technique where the recent conversions are given more weight than the older conversions. The weighting factor is the value of (filter). The formula for the exponential fil...

  • Page 12

    Page 9 function definitions (cont) full scale number (fs) {00001 to +99999} (fs) is for scaling the analog outputs of the indicator. Both the +10v output and the 4-20 ma outputs are automatically scaled by (fs). (fs) = the display value that corresponds to the full scale spans of the outputs, either...

  • Page 13

    Page 10 operating instructions turn-on instructions reference figures 3 and 4 for proper input and output signal wiring. 1. Apply power to the instrument. 2. Verify the instrument readout blanks momentarily, then displays the value of the input signal. Program instructions four front panel switches ...

  • Page 14

    Program instructions (cont) 4. Depress mode (s1) switch once. The text dp followed by the decimal point location will be displayed. Use s2 or s3 to locate the decimal point. To store the selected value in non-volatile memory, depress the mode switch. To exit the mode sequence, without storing the ne...

  • Page 15

    Program instructions (cont) 9. Depress mode (s1) switch once. The text set sp will be displayed momentarily, followed by the current value of setpoint 1. Use s2, s3 and/or s4 to set a new value. To store the selected parameter in non-volatile memory, depress the mode switch. To exit the mode sequenc...

  • Page 16

    Program instructions (cont) 17. Depress mode (s1) switch once. The text sp4 hi or sp4 lo will be displayed. Reference sp1 hi setup instructions on page 11 to set sp4 hi or sp4 lo. 18. Depress mode (s1) switch once. The text set sp will be displayed momentarily, followed by the current value of setpo...

  • Page 17

    Calibration instructions the interface model 9834 can be calibrated using (2) different methods. 1. The first method uses actual or simulated zero and full scale (fs) inputs. Actual inputs may come from a transducer loaded with calibrated weights. Simulated inputs may come from a voltage calibrator....

  • Page 18

    Page 15 calibration instructions (cont) 9. Enable the cal switch. Reference program instructions step 1 on page 10 to enable the cal switch. Note: enabling the cal switch allows only one attempt at calibrating the instrument. If calibration is not successful, repeat the whole calibration procedure. ...

  • Page 19

    Calibration instructions (cont) calibration using the ref switch 1. Attach a calibration adaptor to j1. 2. Apply power to the instrument. 3. The tare and cal led's on the front panel should be illuminated 4. Depress the tare (s2) switch once. The tare led should extinquish. The instrument should now...

  • Page 20

    Page 17 serial rs232 communications general information see figure 4 on page 4 for rs232 connections and wiring information. See figure 5 on page 5 for setting baud rate. The unit is pre-wired for 9600 baud. The data is transmitted in the following format: one start bit; eight data bits; no parity b...

  • Page 21

    Page 18 serial rs232 communications (cont) examples of transmitted data byte 1 2 3 4 5 6 7 8 value in hex 00 01 ff ff fb eb 03 0a value in decimal 0 1 255 255 251 235 3 10 no sp exceeded instantaneous display eot displayed number = -10.45 decimal = xxx.Xx byte 1 2 3 4 5 6 7 8 value in hex 02 42 ff f...

  • Page 22

    Page 19 serial rs232 communications (cont) (m)ode commands to change a mode item, send an 8 character ascii string to the model 9834. The 1st character is always m. The 2nd character identifies the item being changed. The other characters are the item's new value. String transmitted menu item change...

  • Page 23

    Serial rs232 communications (cont) (c)alibrate commands string transmitted menu item changed cc calibrate cu uncalibrate ca x.Xxxxxxe +yy sets the scale factor to x.Xxxxxx * 10 yy examples written in basic. These examples assumed that the rs232 port was opened with buffer # 2. Example 1 print #2, "c...

  • Page 24