Jadever JCN User Manual - Set11
8
Use the numeric key to set the format directly.
12)
SET11
filtering setting: 1,2,3,4
Set the filtering level in which the stable indication turns on .The higher the setting, the
slower stabilization time.
13)
SET12
Precondition for
zeroing or taring setting: stable, auto, always
stable :
T
aring or zeroing can be not action without stable indication after pressing the key
TARE
or
ZERO
always:Taring or zeroing can be action without stable indication after pressing the key
TARE
or
ZERO
auto: Press key
TARE
or
ZERO
, but it works only when get the stable indication.
14)
SET13
weight memory parameter: on,off
On: display the last weight when powering on again
Off: do not display the last weight when powering on again
15)
SET14
the kinds of checking: Qty.ck,Wt.ck
Qty.ck: start amount checking
Wt.ck: start weighing checking
16)
SET15
parameter initialization
press key
G N/SET
then key
M+/SAVE
to start initialization. It is finished when
displaying OK.
6. Single and Multi-point Calibration
Note: Before calibration, please set the capacity first. The unit used in calibration is the one
that has been set before. During the calibration procedure, LONG press
ZERO/ESC
to
return to normal weighing mode without saving.
Here we take 3kg/0.2g as an example
1. Press and hold
TARE/
while powering on.
Do not release it till the window displays
―
CAL”
2. With no load on the weighing pan,
press
TARE/
to start zero point calibration.
3. Wait till the window displays the first calibration value.
Note: The first calibration value is default. If the value
has been changed, the default value is 1/3 of full load.