Keithley 4200-SCS Startup Manual - Startup Guide
Startup Guide
3
Step 3
Connect DUT test fixture to 4200-SCS
Figure 3
shows how to connect a test fixture for a 4-terminal device to the 4200-SCS. The test fixture must
be equipped with standard 3-lug female triax connectors.
• Connections to SMUs – To connect directly to an SMU, use a Mini triax cable that is terminated with a
miniature male triax connector on one end and a standard 3-slot male triax connector on the other end.
• Connections to PreAmps – To connect to a PreAmp, use a cable that is terminated with a standard
3-slot male triax connector on both ends.
NOTE
Connections are snug; push firmly.
Figure 3: DUT test fixture connections to 4200-SCS
Step 4
Power-up & log-in
A. Make sure the power switch is in the O (out) position. The POWER switch is located on the front panel
in lower right-hand corner.
B. Plug the male end of the line cord into a properly grounded AC line power receptacle.
C. Turn on the 4200-SCS by pushing in the
POWER
switch to the I (in) position.
D. When prompted, simultaneously press
Ctrl - Alt - Del
.
E. At the KIUSER prompt, press
ENTER
. There is no password for this account.
CAUTION When first starting a KTE-Interactive software tool, you must
answer “Yes” to an on-screen license agreement. Answering
“No” makes your system nonfunctional until you reinstall the
software.
INSTR
UMENT
CONNECTIONS
SMU ONL
Y
SMU AND GNDU
GNDU
INSTR
UMENTS
SLO
T
8
SLO
T
7
SLO
T
6
SLO
T
5
SLO
T
4
SLO
T
3
SLO
T
2
SLO
T
1
S
E
N
S
E
F
O
R
C
E
C
O
M
M
O
N
SENSE LO
GU
SENSE LO
COMMON
COMMON
FORCE
SENSE
GU
SENSE LO
SENSE
FORCE
SENSE LO
SENSE
FORCE
SENSE LO
SENSE
FORCE
SENSE LO
SENSE
FORCE
PA CN
PA-1 REMO
T
PA-1 REMO
T
PA-1 REMO
P
RE
MP
4200-P
A-1 REMO
CONTR
OL
OUT
IN
INSTR
UMEN
T
CONNECTIONS
SMU ONL
Y
SMU AND GNDU
GNDU
INSTR
UMENTS
SLO
T
8
SLO
T
7
SLO
T
6
SLO
T
5
SLO
T
4
SLO
T
3
SLO
T
2
SLO
T
1
S
E
N
S
E
F
O
R
C
E
C
O
M
M
O
N
SENSE LO
GU
SENSE LO
COMMON
COMMON
FORCE
SENSE
GU
SENSE LO
SENSE
FORCE
SENSE LO
SENSE
FORCE
PA CNTRL
SENSE LO
SENSE
FORCE
PA CNTRL
SENSE LO
SENSE
FORCE
PA CNTRL
OUT
IN
PA CNTRL
Four
Terminal
Device
Four
Terminal
Device
“FORCE”
Connectors
“FORCE”
Connectors
on SMUs
PreAmps
“FORCE” Connector
on GNDU
“FORCE” Connector on GNDU
Test Fixture
Test Fixture
Connections to SMUs
Connections to SMUs/PreAmps