Labomed CxL User Manual - page 26
Trouble
Cause
Remedy
9. Objective hits the specimen when
The specimen is upside down
Set the specimen correctly with the
an objective is switched to a higher
cover glass on the top
magnification objective
The cover glass is too thick
Use a cover glass with thickness of
0.17mm
The stage is raised too high
Lower the stage
The slide is slipped from the slide
Re-position the slide in the slide
holder
holder
Slide is of excessive thickness
Use slide having thickness between
0.9 and 1.4mm
10. Bulb/LED does not light
Bulb/LED is not mounted
Attach a bulb/LED
Bulb/LED is blown
Replace the bulb/LED
The power cord is unplugged
Plug it securely
Fuse is blown
Check and replace with live fuse
Battery is low
Charge battery
11. Bulb/LED blows easily
The specified bulb/LED is not used
Replace with a specified bulb/LED
23