OHAUS 4000W Series Instruction Manual - Valor
Valor
™ 4000W Series EN-9
4. MENU SETTINGS
The User Menu allows the customizing of scale settings.
4.1 Menu Navigation
User Menu:
Menu:
C.A.L
S.E.t.U.P
r.E.A.d
M.O.d.E
U.n.i.t
L.O.C.k
E.n.d
Menu
Items:
Span
Lin
GEO
End
Reset
Pwr.Un
A.Tare
Ir.Func
Ir.Adj
Accum
End
Reset
Stable
Filter
AZT
Light
Sleep
A.Off
End
Reset
Percnt
Check
End
kg
g
oz
lb
lb:oz
End
Reset
L.Cal
L.Setup
L.Read
L.Mode
L.Unit
End
Notes:
Some Units/Modes may not be available in all models.
When LEGAL FOR TRADE is set to ON (see section 4.9), the menu settings are
affected.
To Enter the Menu Mode
Press and hold Menu until
[mMeNU] (Menu) is displayed. When released the first sub-
menu
[C.a.l] (Cal) will be shown.
Press Yes to enter the displayed sub-menu or press No to advance to the next.
Selecting a sub-menu will display the first menu item. Press Yes to view the menu
item setting or press No to move to the next menu item. When viewing the setting,
press Yes to accept the setting, or press No to change the setting. When
[End] is
displayed, press Yes to return to the sub-menu selections or No to return to the first
item in the current menu. Bold indicates factory default setting.
Note:
The Calibration / Menu Mode indicator is displayed when in the Menu Mode.
4.2 Cal Menu
Enter this menu to perform calibrations.
Span
[SpaN]
(yes, no)
Initiates a span calibration procedure (zero and span).
Lin
[LIN]
(yes, no)
Initiates a linearity calibration procedure (zero, mid-point and span).
GEO
[GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration
based on the current location. Settings from 0 to 31 are available with 12
being the default. Refer to table 4-1 to determine the GEO factor that
corresponds to your location.
End Cal [End]