    ™ pioneer balances instruction manual.

  • Page 2: 1. Introduction

    En-1 pioneer ™ 1. Introduction 1.1 safety precautions please follow these safety precautions: • verify that the ac adapter input voltage matches the local ac power supply. • use the balance only in dry locations. • do not operate the balance in hostile environments. • do not drop loads on the platfo...

  • Page 3

    Pioneer ™ en-2 2.2 install components 0.1g and 0.01g models 1) install wind-ring (incal only) 2) install pan support 3) install pan 0.001g and 0.0001g models 1) install side doors - insert fully into top frame then down over retainer. 2) install panels - insert bottom edge in groove then press until...

  • Page 4: 3. Operation

    En-3 pioneer ™ 2.3 level balance level the balance on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat sources or rapid temperature changes. 2.4 connect power for use with csa certified (or equivalent approved) power source, which must have a limited circuit outpu...

  • Page 5: 4. Settings

    Pioneer ™ en-4 accuracy of the count. The balance automatically recalculates the average piece weight when the number of parts on the pan is less than three times the original sample size. The display shows [ apww.Opt] each time the apw is optimized. 3.3 percent mode – use this mode to measure the w...

  • Page 6

    En-5 pioneer ™ 4.2 calibration menu [ .Cal. ] incal or span calibration should be performed daily and when the room temperature changes. Incal [ incal] calibrates the balance using an internal mass. Span calibration [ span] uses two weight values: zero and a weight between 50% and 100% of the capaci...

  • Page 7

    Pioneer ™ en-6 lsd – the least significant digit (lsd) is the number of displayed divisions (d) by which the weight is incremented. The values are 1d, 2d, 5d, 10d, 100d or 0.5d. Press no to go to the next setting, press yes to accept. (example lsd = 1d) note: lsd options are limited if the readabili...

  • Page 8: 5.0 Maintenance

    En-7 pioneer ™ 5.0 maintenance 5.1 troubleshooting symptom possible cause remedy cannot turn on no power to balance verify connections and voltage. Poor accuracy improper calibration unstable environment perform calibration move balance to suitable location cannot calibrate unstable environment inco...

  • Page 9: 6. Technical Data

    Pioneer ™ en-8 6. Technical data ambient conditions – the technical data is valid under the following ambient conditions: ambient temperature: 10°c to 30°c relative humidity: 15 % to 80 % at 31°c non-condensing, decreasing linearly to 50% at 40°c height above sea level: up to 2000 m operability is a...

  • Page 10

    En-9 pioneer ™ 6.3 communication 6.3.1 commands the rs232 interface allows a computer to control the balance as well as receiving data such as the displayed weight. The balance will return “es” for invalid commands. Command function ip immediate print of displayed weight. P print displayed weight (u...

  • Page 11

    Pioneer ™ en-10 disposal in conformance with the european directive 2002/96 ec on waste electrical and electronic equipment (weee) this device may not be disposed of in domestic waste. This also applies to countries outside the eu, per their specific requirements. Please dispose of this product in a...