OHAUS Pioneer Instruction Manual - 6. Technical Data
Pioneer
™
EN-8
6. TECHNICAL DATA
Ambient conditions – The technical data is valid under the following ambient conditions:
Ambient temperature: 10°C to 30°C
Relative humidity: 15 % to 80 % at 31°C non-condensing, decreasing linearly to 50% at 40°C
Height above sea level: Up to 2000 m
Operability is assured at ambient temperatures between 5°C and 40°C
Power – AC adapter, Balance power input 8-14.5 VAC, 50/60Hz 4VA or 8-20 VDC, 4W
Protection – Protected against dust and water, Pollution degree: 2, Installation category: Class II
6.1 Drawings
0.01g and 0.1g Models
0.0001g and 0.001g Models
6.2 Specifications
*C= Internal calibration
MODEL
PA64* PA114* PA214* PA213* PA413* PA512* PA2102* PA4102* PA4101*
Capacity (g)
65
110
210
210
410
510
2100
4100
4100
Readability (g)
0.0001
0.001
0.01
0.1
Repeatability (sd)
0.0001 0.001
0.01 0.1
Linearity (g)
±0.0002
±0.002
±0.02
±0.1
Tare Range
To capacity by subtraction
Stabilization 3
seconds
Span cal mass (g)
50 or
60
50 or
100
100 or
200
100 or
200
200 or
400
200 or
500
1000 or
2000
2000 or
4000
2000 or
4000
Lin. cal masses (g)
20,
50
50,
100
100,
200
100,
200
200,
400
200,
500
1000,
2000
2000,
4000
2000,
4000
Pan size (in / cm)
3.5 / 9 dia.
4.7 / 12 dia.
7.1 / 18 dia.
Net Weight (lb/kg)
10.1 / 4.6 (*11.2 / 5.2)
10 / 4.5 (*11.1 / 5.1)