6.3 Communication
6.3.1 Commands
The RS232 Interface allows a computer to control the balance as well as receiving data such as the
displayed weight. The balance will return “ES” for invalid commands.
Command Function
IP
Immediate Print of displayed weight.
P
Print displayed weight (uses Stable ON/OFF menu settings).
CP
Continuous Print.
xP
Interval Print x = Print Interval (1-3600 sec)
T
Same as pressing Zero Key.
ON
Turns balance ON.
OFF
Turns balance OFF.
PSN
Show Serial Number.
PV
Version: Print product name, software revision and LFT ON (if LFT is set ON).
PU
Print current mode/unit
x#
Set PC ref wt (x) in grams
P#
Print PC ref wt
x%
Set % ref wt (x) in grams
P%
Print % ref wt
CTS
RTS
5
9
1
6
RxD
TxD
GND
6.3.2 RS232 (DB9) Pin Connections
Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
6.4 Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark
Standard
This product conforms to the EMC Directive
2004/108/EC
, the Low Voltage Directive
2006/95/EC and the Non-Automatic Weighing Directive 90/384/EEC. The complete Declaration
of Conformity is available from Ohaus Corporation.
AS/NZS4251.1 Emission; AS/NZS4252.1 Immunity
CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1
ISO 9001 Registration In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by
Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant
with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to the
ISO 9001:2000 standard.