QUATTRO EMS GM380E Operation Manual - page 6
THE DEVICE CONTROLS
Intensity
The intensity knobs are located on the front of the unit for the strength adjustment
of the stimulation and also function as ON/OFF controls.
Mode
The Mode key is used to select / set the type of treatment utilized. The available
stimulation modes for each unit are as indicated in the following table:
Identification
Function as
Available mode
E EMS
S (Synchronous), C (constant),
A (Alternation)
TIME
Treatment Time of device can be pre-select / set with Time key. There are seven
programs fixed duration of 5, 10, 15, 20, 30, 45, 60 minutes and continue. Press
the key until engaged in position desired.
5 min : 1/4 circle icon flash.
10 min : 2/4 circle icons flash.
15 min : 3/4 circle icons flash.
20 min : 4/4 circle icons flash.
30 min : 2/4 circle icons light on.
45 min : 3/4 circle icons light on
60 min : 4/4 circle icons light on.
Continue : the timer icon become empty.
SET/ENT
The SET/ENT keys are used to select the type of treatment utilized. The type of
treatment utilized includes cycle On time, On Ramp time, cycle Off time and Pulse
Rate. The selected parameter will flash during adjust.
▲▼
The
▲▼
keys regulate the number of On, Ramp, Off and Pulse Rate value.
DANGER
The device does not have AP/APG protection. Explosion hazard is possible if used
in the presence of explosives, flammable materials or flammable anesthetics.
Caution should be used when applying the device to patients suspected of having
heart disease. Further clinical data is needed to show if there are adverse side
effects on those with heart disease.
ATTACHING THE LEAD WIRES
-
4
-