Quickie puma 20 Technical Manual - page 184
Puma 20/40
184 | Trouble shooting
Trip Text
Trip
Code
Description
User Switch detached
-
The Omni has detected that the User Switch has become disconnected.
• Check all cables and connectors between the Omni and the User Switch.
If the trip is still present after the above checks have been made, then the User Switch may be
defective. Contact your service agent.
If it is required to use the Omni without a User Switch being connected, then the parameter,
Switch Detect, should be set to Off. If a User Switch is not used the responsibility for that
decision lies with the healthcare professional.
Gone to Sleep
This occurs when the control system has been left inactive for a time greater than the
parameter Sleep Timer.
An entry is made in the system log each time this occurs.
Charging
This occurs when the control system detects that a charger is connected to either Inhibit 1
or Inhibit 3. Refer to section 2.3 for connection details The Battery charging screen will be
displayed during charger connection.
An entry is made in the system log each time this occurs.
If an On-Board Charger is used:
• Disconnect the charger from the AC supply.
If an Off-Board Charger is used:
• Disconnect the charger from the Wheelchair.
If the trip is still present after the charger has been disconnected then the
Joystick Module may be defective.