Quickie puma 20 Technical Manual - page 188
Puma 20/40
188 | Trouble shooting
Flash
Code
Fault
source
Sub
code
(HHP)
Meaning
09, 10
DX BUS
All
DX BUS hardware fault
• Check DX BUS cables
• Turn the system off and on several times. Make sure to wait 10
seconds before you turn the system on again.
• If all DX BUS connections are OK and turning the system off and on does not help, contact
the Service Department of Sunrise Medical HCM
All
other
Internal fault
All
Contact the Service Department of Sunrise Medical HCM
DX2 System condition indications
Condition
DX2-ACT
indication
DX2-REM550
indication (DX2)
DX-REMG90
indication (DX)
DX System Faults
CLAM is enabled
None, but ACT will
not
drive actuators
None
None
Master Remote in Fault State
FC6
FC according to
fault
FC according to
fault
DX2-ACT detects CAN Fault
FC6/9/10,
depending
on actual fault
FC according to
fault detected by
Remote
FC according to
fault detected by
Remote
Actuator System
Configuration Faults
Module ID Collision
(2 modules with same ID)
FC6, No DX BUS
communication
Actuator system will
be "missing"*
None
ACT Module Missing "Slow1/Slow2"
None
FC2, Drive inhibit*
Not detected
ACT Module Missing "Stop"
None
FC2*
Not detected
Actuator Module Local Fault
FC2 on remaining
ACT Module
Drive Inhibit
Drive Inhibit
Actuator Module Local Fault
See DX2-ACT Flash
Code List
None
None
ACI “Slow” function
active
ACI "Flash Code 2"
FC2
FC2
FC2
ACI "Stop (Drive Inhibit)"
FC2
Drive inhibit
Drive inhibit
ACI "System Slowdown"
FC2
FC2
FC2
ACI "Local Slow"
FC2
Reduced speed on
Speedometer
None
ACI in Speed Pot or Speed Limit
mode, with outputs less than 100%
FC2
Reduced speed on
Speedometer
None
*The REM550 responds to a missing DX2 actuator system if the following parameters are set:
Parameter
Required Value
Actuator System Type
DX2 Actuator System
Actuator System Is Critical Yes
Actuator System Missing "Stop Drive" or "Slowdown Drive", as desired.