Rixen M70-D Instruction Manual - page 5
needs to adjust the instrument, please operate the instrument as the steps below
F-1. When the power is off, hold firstly, and then press to turn on,「CAL」will appear
on the display left below and the instrument will make a voice “Be” every 4 seconds, it
represents the instrument is in the adjustment model.
F-2. Please use the small type screw driver to do fine-tuning on “CAL” at the right side of the
instrument to display the value and get the calibration value.
F-3. When calibration is finished, please turn off the power to store data.
◆Before calibrating, please check if the power is enough and then process the calibration
◆If you don’t know if the instrument needs to be adjusted, please contact our technical
department firstly and don’t change any original setting.
◆The built-in calibration value of this instrument is 18.4
%
±1
%
。
G. Power inspection function
The display left above has a power indicator symbol, if the power is not enough, the symbol will
flash and make a voice “Be” warning by fits and starts. Please replace the new batteries immediately
and then operate the instrument
5. Signal Output
This instrument has synchronal output function, and it can be used to connect the related
peripheral device like a recorder, controller, PC Data-Logger…..etc.
※output signal:DC 0 ~ 1V,10mV∕
%
6. Care and Maintenance
1. This instrument has the accurate temperature compensation function, but, if it is used in the place
with high temperature or high humidity, it will produce deference or error message.
2. The sensor (needles) is the electric conduction elements of connecting the measured material. In
order to avoid damage to the instrument and electric shock, when the measured material has
voltage or static electricity, please clear the voltage or cut off the power and then go to measure.
3. Please use the two needles to pierce into the material at the same time and keep them for a while
to get the more accurate value.
4. For the hard material or the material need deep inspection, please change to use the special needle
(option)
5. To avoid the incorrect reading, if the low symbol appears, please replace the battery before using
6. The sensor (needles) is the main element, if the sensor has damage or other problems, please
replace the new one to use continually.
7. When not in use, please remove the battery from the instrument, put the main instrument and all
accessories into the protection case and keep the kit in a dry, dust-free environment out of direct
sunlight.
8. Please contact the local agent or our customer service department for any question or product
repair.