Rixen TX-600N Instruction Manual

Summary of TX-600N

  • Page 1

    Contents tx-600n specifications-----------------------------------1 instrument descriptions ----------------------------------2 display descriptions --------------------------------------3 key descriptions ------------------------------------------4 instructions -------------------------------------...

  • Page 2: ■Tx-600N Specifications

    ■tx-600n specifications input types k, j, e, t, b, r, n, s, c thermocouple types; with dual-channel input measurement range k: -200.0 ~ +1370.0℃(-328.0~+2498.0℉) j:-200.0 ~ +1200.0℃(-328.0~+2192.0℉) e:-210.0 ~ +1000.0℃(-346.0~+1832.0℉) t:-220.0 ~ +400.0℃(-364.0~+752.0℉) b:+320.0 ~ +1800.0℃(+608.0~+3...

  • Page 3: Instrument Descriptions

    Instrument descriptions 部名稱 ○ 1 multi-function lcd display ○ 8 battery cover ○ 2 function key and setting key ○ 9 ac adaptor input ○ 3 record key and thermocouple switch key ○ 10 usb output port ○ 4 movement key ○ 11 rs-232 output connector ○ 5 movement key ○ 12 t1 probe input connector ○ 6 power sw...

  • Page 4: Display Descriptions

    Display descriptions display descriptions battery power symbol manual shut down buzzer on cal under calibrating hold data hold rcd data-logging max maximum value min minimum value avg average value t1 t1 probe t2 t2 probe hi.A high point alarm lo.A low point alarm english words zone thermocouple typ...

  • Page 5: Key Descriptions

    Key descriptions key function descriptions lcd display on/off press one second (be-) can turn on the power. Press one second (be---) can turn off the power. If the instrument is left without any operation for five minutes, it will turn off automatically. + + manual turn off when the device is on off...

  • Page 6

    Rcd recode mode under testing mode, and there is no recording data, press to enter rcd record mode. The will flashes when record. Press or to stop recording. Rcd reading mode under testing mode, press to into the data reading hold mode. Use to switch the displayed record value; long press the can sw...

  • Page 7: ■Instructions

    ■instructions a. Testing mode: press and to switch display conditions. In the testing mode, the display content as below: status display contents normal conditions probe condition, probe type, warning data condition, power indication t1 real-time temperature display area a: t1 measuring data value a...

  • Page 8: B.Rcd Record Mode:

    B.Rcd record mode: ※please use the ac adaptor when you need to record for a long time under testing mode, press into the record mode, it will show the sampling rate before starting record. Under the record mode press and can switch the display, the will flashed when recording. T1 and t2 will show th...

  • Page 9: D.Hold Mode:

    D.Hold mode: in lock mode, display content as description below: e. Rcd reading mode: under the testing mode, press to enter the rcd clear mode. Use and to switch the displayed record value. Press the again can back to the testing mode. Press and hold the and can speed us the switch. Press and hold ...

  • Page 10: F. Rcd Clear Mode:

    F. Rcd clear mode: under the rcd reading mode or the hold mode, press and hold to enter the rcd clear mode. Then will be displayed. Press and to select and use to confirm. The twinkled means not to clear the data, and the twinkled means to clear the data. When clearing the data, the will twinkled, p...

  • Page 11: Setting Mode Instructions

    Setting mode instructions 1. Alarm setting mode this mode is for setting the hi.A and lo.A, press to change the digit you want to change, and press and to select the circulation figures from 0 to 9. It’s thousands of bits can be switched to the negative sign. Press to confirm the setting. Setting ra...

  • Page 12

    ■ rs-232 transmission agreements ※please connect the ac adaptor if it required a long time to transfer the data. Rs-232 is for one-way data transfer, receive and input the signal by three grounded wires. Recommending using the transmission line which is manufacture by oe factory or shorter than 10 m...

  • Page 13

    C. Request data frame ex: read the data from address 02h (read 5-byte of data from address 0002h) slave address function starting address hi starting address lo no. Of word hi no. Of word lo crc lo crc hi 03h 03h 00h 02h 00h 05h 25h ebh response data frame ex. Response data =10-byte slave address fu...

  • Page 14

    【example】set mode: 1. K type thermocouple 2. Temperature unit=℃ 3. T1 no error 4. T2 no error 5. Power74~50% mod setting mode 0= k type thermocouple, 1= j type thermocouple, 2= e type thermocouple, 3= t type thermocouple, 4= b type thermocouple, 5=r type thermocouple, 6=n type thermocouple, 7= s typ...

  • Page 15: ■ Precautions

    D. Cable illustrated ■ precautions 1. This instrument has a waterproof function; please do not use it in a high temperature environment or with corrosive materials to avoid leakage or damage. 2. Proposed to use the company's original signal cable (sold separately) to avoid the instrument inability t...

  • Page 16: Memo

    Memo ------------------------------------------------------------------ ------------------------------------------------------------------- ------------------------------------------------------------------- ------------------------------------------------------------------- ------------------------...