Rixen TX-600N Instruction Manual - ■Tx-600N Specifications
■TX-600N Specifications
Input Types
K, J, E, T, B, R, N, S, C thermocouple types; With Dual-channel input
Measurement Range
K: -200.0 ~ +1370.0℃(-328.0~+2498.0℉)
J:-200.0 ~ +1200.0℃(-328.0~+2192.0℉)
E:-210.0 ~ +1000.0℃(-346.0~+1832.0℉)
T:-220.0 ~ +400.0℃(-364.0~+752.0℉)
B:+320.0 ~ +1800.0℃(+608.0~+3272.0℉)
R:-20.0 ~ +1700.0℃(-4.0~+3092.0℉)
N:-200.0 ~ +1300.0℃(-328.0~+2372.0℉)
S:-20.0 ~ +1750.0℃(-4.0~+3182.0℉)
C:0.0 ~ +2300.0℃(+32.0~+4172.0℉)
Accuracy
± 0.1% of rdg + 0.1℃ (In 23±5℃ operating environment)
± 0.1% of rdg + 0.2℃(Out off 23±5℃ operating environment)
※Built-in cold junction compensation function
Resolution
K, J, E, T, N, C Type: 0.1℃/0.1℉;
B, R, S Type: 0.5℃/0.5℉
Reading Rate
Approx 0.4 sec
Logging Sampling Rate
2 second to 120 minutes (User Selectable)
Memory
16,000 readings x 2CH
Baud Rate
57,600
Main Functions
Switchable 9 kinds of thermocouple type input, Hi/Lo Alarm, T1-T2,
Data hold, Max/Min/Avg Functions, USB/RS-232 interface, Perpetual
calendar, Data-Logging, Switchable ℃/℉ , AC/DC power, Battery
sign and low battery warning, Auto / Manual shutdown, Calibration
function, Large LED back-light, IP66 water and dust proof.
Output
Software with USB Interface cable, RS-232 Output
Power Source
One 9 V battery or AC Adaptor
Dimensions / Weight
150 x 75 x 28 mm, Approx 320g (battery included)
Input Connections
Standard mini thermocouple socket x 2
Operating Environment
-20~+60℃; 0~100%RH
※About accuracy and sampling rate excludes errors generated by temperature probe.
Option Accessories
Temperature Probes LP series
Please select from LP series temperature probe (Customized)
TU-RS232-W
RS-232 interface cable
TU-USB-W
USB interface cable and WINDOWS software
TU-609
9 V battery
※Specifications are subject to change without notice.