— PRELIMINARY —
GILIAN GILIBRATOR 2 CALIBRATION SYSTEM
20
Sensidyne Document No. 850190M (Rev G)
3.2.2
Cabling Connections
IMPORTANT
Turn off the unit before connecting any cables.
If you are using a Piston Cell Assembly, insert the 15-
Pin Interface Plug directly into the Connecting Jack lo-
cated on the back of the Piston Cell (refer to Figure
2.3). Be sure to properly match up the connectors be-
fore engaging.
If you are installing a Wet Cell Assembly on a 15-Pin
Control Base, insert the 15-Pin Interface Plug from the
Control Base into the 15-to-9 pin adapter. Insert the
adapter into the Connecting Jack located on the back of
the Wet Cell (refer to Figure 2.4). Be sure to properly
match up the connectors before engaging.
If you are installing a Wet Cell on a 9-Pin Control Base,
insert the 9-Pin Interface Plug from the Control Base
directly into the Connecting Jack located on the back of
the Wet Cell (refer to figure 2.4). Be sure to properly
match up the connectors before engaging.
Connect the appropriate optional printer (or computer)
cable to the Printer Jack located on the left side of the
Control Unit. Make certain the cable and Printer Jack
connectors match up before making a complete con-
nection. Once the cable has been properly connected,
turn on the printer (or computer) and make certain it is
prepared to receive data before turning on the Gilibra-
tor 2 unit.
3.2.3
Adding Soap Solution (Wet Cell only)
WARNING
Only use Soap Solution Part No. 800450 with the Flow
Cell Assembly. Use of other soap solutions can damage
the Flow Cell.
Add soap solution to the unit as follows:
1)
Remove the Storage Tubing from the Air Outlet
Boss (see Figure 2.3) of the Flow Cell. Fill the dis-
penser bottle provided with Gilibrator soap solu-
tion. Using the rubber Storage Tubing as a funnel,
slowly add soap solution from the dispenser (refer
to Figure 3.4).
2)
The amount of soap needed is determined by de-
pressing the Bubble Initiate Button and holding it
down in the lower position. Continue to add
enough soap solution until the angled edge at the
bottom of the Bubble Generator Ring is immersed
in the solution.
CAUTION
Do Not Overfill.
3)
After filling is completed, the rubber Storage Tub-
ing may be removed completely. Recap the soap
dispenser bottle for later use.
NOTE
If the Wet Cell Assembly is not going to be used for an
extended period of time, reinstall the rubber Storage
Tubing between the inlet and outlet bosses. This pre-
vents the solution from evaporating and altering the soap
concentration in the solution.
WARNING
If you are shipping the Wet Cell, you must remove the
rubber Storage Tubbing between the inlet and outlet
bosses.
3.2.4
Sampling Source Connection
Connect the air sampler to be calibrated to the Air Out-
let Boss of the Wet Cell Assembly or the Inlet Tube on
the Piston Cell Assembly. Use 1/4” ID tubing in either
case.
WET CELL NOTE
A liquid trap between the sampler and the Wet Cell
Assembly is recommended. This prevents moisture from
carrying over into the sampler during continuous cali-
bration periods.