Operating instructions vegadis 176 external indicating and adjustment display without external energy for front panel mounting document id: 47916.
2 contents vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 contents 1 about this document 1.1 function ..............................................................................................................................
3 contents vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 safety instructions for ex areas takenoteoftheexspecificsafetyinstructionsforexapplications. These instructions are attached as documents to each instrument with ex ...
4 1 about this document vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 1 about this document 1.1 function this operating instructions manual provides all the information you need for mounting, connection and setup of the inst...
5 2 for your safety vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 2 for your safety 2.1 authorised personnel all operations described in this operating instructions manual must be carried out only by trained specialist per...
6 2 for your safety vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 2.5 safety label on the instrument the safety approval markings and safety tips on the device must be observed. 2.6 eu conformity thedevicefulfilsthelegalrequ...
7 3 product description vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 3 product description 3.1 configuration the scope of delivery encompasses: • indication and adjustment display vegadis 176 • mounting material • documen...
8 3 product description vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 3.4 packaging, transport and storage your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is ...
9 4 mounting vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 4 mounting 4.1 mounting location, installation position the instrument is designed for use in a front panel. The installation position is horizontal. 4.2 mounting ...
10 5 connect to the signal circuit vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 5 connect to the signal circuit 5.1 preparing the connection always keep in mind the following safety instructions: • connect only in the compl...
11 5 connect to the signal circuit vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 1 3 2 4 5 fig. 5: terminal assignment vegadis 176 terminal function polarity notes 1 output, sensor + display with backlight connection activ...
12 5 connect to the signal circuit vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 1 2 + – – + 1 3 2 4 5 fig. 8: wiring plan vegadis 176 to passive sensors, with backlight 1 to the sensor 2 to power supply or processing system...
13 5 connect to the signal circuit vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 5 1 2 + ( ) (-) 6 7 8 4...20ma 2 1 6 4 3 5 1 3 2 4 5 + - 1 2 3 4 18 17 16 15 1413 1211 10 9 8 7 6 5 n l - + + - on 1 2 1 2 3 4 % fig. 12: con...
14 6 setup vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 6 setup 6.1 indication and adjustment 1 2 3 4 7 6 5 fig. 13: display and adjustment elements of vegadis 176 1 symbol: adjustment menu locked 2 symbol: error 3 symbol: ...
15 6 setup vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 set parameter in the 7-segment display. Press the "ok" key to edit and then enter the user code. Text parameters iftheadjustmentpositioncontainstextparameters,atfirs...
16 6 setup vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 example: sc__20 = 100.0 means display 100.0 with measurement current 20 ma. The unit selected under unit is used in the display. Value range: -19 999 … 99 999 in this ...
17 6 setup vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 menu submenu menu item default values setup - decim 1 dec - sc__4 0.0 - sc__20 100.0 - unit % - text - exprt systm ucode 0000 frset no input decim 1 dec sc__4 0.0 sc...
18 7maintenanceandfaultrectification vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 7 maintenanceandfaultrectification 7.1 maintenance if the device is used properly, no special maintenance is required in normal operation. 7....
19 7maintenanceandfaultrectification vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 diagnosic code short text rectificationmeasure f100 sensor error • check electrical wiring • check sensor • check sensor parameter adjustme...
20 8 dismount vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 8 dismount 8.1 dismounting steps take note of chapters "mounting" and "connecting to power supply" and carry out the listed steps in reverse order. 8.2 disposal the...
21 9 supplement vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 9 supplement 9.1 technical data note for approved instruments thetechnicaldataintherespectivesafetyinstructionsarevalidforapprovedinstruments(e.G.With exapprova...
22 9 supplement vegadis 176 • external indicating and adjustment display without external energy for front panel mounting 47916-en-170801 current range 3.6 … 22 ma overcurrent resistance 200ma(fuseprotection,supplyside) hart signal notinfluenced(hart-transparent) reverse voltage protection available...
23 9 supplement vegadis 176 • external indicating and adjustment display without external energy for front panel mount- ing 47916-en-170801 9.3 industrial property rights vega product lines are global protected by industrial property rights. Further information see www.Vega.Com. Only in u.S.A.: furt...
Printing date: vega grieshaber kg am hohenstein 113 77761 schiltach germany 47916-en-170801 all statements concerning scope of delivery, application, practical use and operat- ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subjec...