7Maintenanceandfaultrectification
VEGADIS 176 • External indicating and adjustment display without external energy for front panel mount-
ing
47916-EN-170801
Diagnosic code Short text
Rectificationmeasure
F100
Sensor error
•
Check electrical wiring
•
Check sensor
•
Check sensor parameter adjustment
S901
Input signal too
low
•
Check sensor output for defects and
deviation from normal character-
istics
•
Check sensor parameter adjustment
S902
Input signal too
large
The instrument has diagnostic functions for its own electronics. If an
error is detected, a diagnostic message is displayed.
Diagnosic code Short text
Rectificationmeasure
F261
Electronics mod-
ule
Replace electronics
F283
Memory content
•
Restart instrument
•
Carry out device reset
•
Replace electronics
F431
Factory calibration Replace electronics
Diagnosic code Short text
Rectificationmeasure
M561
Display exceeded Check scaling
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
Thehotlineismanned7daysaweekround-the-clock.Sinceweoffer
this service worldwide, the support is only available in the English
language. The service is free, only standard call charges are incurred.
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
7.3 How to proceed if a repair is necessary
Youcanfindaninstrumentreturnformaswellasdetailedinforma-
tion about the procedure in the download area of our homepage:
www.vega.com.
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
•
Printandfilloutoneformperinstrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Please contact the agency serving you to get the address for
thereturnshipment.Youcanfindtheagencyonourhomepage
www.vega.com.
Electronics diagnosis
Configurationdiagnosis
24 hour service hotline
Reaction after fault recti-
fication