5 Setup
Electronics module • VEGAFLEX 80 series
43656-EN-170531
Thenallapplication-specificsettingsmustbeenteredagain.Seethe
following chapter "Setup steps"
Firstofall,youhavetotransferthedevice-specificsensordatatothe
new electronics module.
Under "Instrument search (serial number)" you can download the
specificsensordataasXMLfileandthePDFfile(onlywithSILquali-
fiedinstruments)directlyintothesensor.
If you can access the sensor via the adjustment software PACTware
and a DTM, a direct transmission is the easiest and quickest method.
For this you need an Internet access.
1. Start PACTware and connect to the corresponding sensor.
2. Openthefield"Maintenance" and select"Electronics exchange"
3. After the connection is established, you can choose how the sen-
sor data should be loaded into the sensor.
Choose "Load sensor data from the Internet"
4. Enter in the next step the serial number of your sensor.
Youcanfindtheserialnumberoutsideonthetypelabelofthe
instrument and inside the housing.
5. As soon as the sensor data are ready, you can load the data
directly into the sensor.
This procedure lasts approx. 5 minutes.
6. Thenallapplication-specificsettingsmustbeenteredagain.See
the following chapter "Setup steps"
InstrumentswithSILqualification
After the transfer of the sensor data, a check sum is displayed. You
have to verify the correct transmission by means of a check sum. Only
thenistheinstrumentreadyforoperationagainandSILqualified.
The PDF document "SIL electronics exchange check sum certificate"
is opened automatically.
After that, a list of all safety-relevant data is displayed.
Compare the displayed check sum with the data of the PDF docu-
ment.
If the check sums match, the transmission was successful. Record
the comparison of the check sums.
Firstyouhavetotransferthedevice-specificsensordatatothenew
electronics module.
Youcandownloadtheseindividual,device-specificdataofyoursen-
sor from our homepage.
This method is recommended if you have no Internet access on site.
1. Go via our homepage "www.vega.com" to "Instrument search
(serial number)".
2. Enter the serial number of your sensor.
Sensor data - on site via
PACTware
Sensor data - on site via