Vega VEGAPULS WL 61 Quick Setup Manual - 7 Supplement
19
7 Supplement
VEGAPULS WL 61 • Foundation Fieldbus
47097-EN-170126
7 Supplement
7.1 Technical data
Note for approved instruments
The technical data in the respective safety instructions are valid for approved instruments (e.g. with
Exapproval).Thesedatacandifferfromthedatalistedherein,forexampleregardingtheprocess
conditions or the voltage supply.
Electromechanical data - version IP 66/IP 68 (2 bar)
Cable entry
IP 68 cable gland
Connection cable
Ʋ Configuration
two wires, one Kevlar cable, braiding, cover
Ʋ Standard length
6 m (19.69 ft)
Ʋ Max.length
550 m (1804 ft)
Ʋ Min. bending radius
25 mm (0.984 in) with 25 °C (77 °F)
Ʋ Diameterapprox.
8 mm (0.315 in)
Ʋ Wire isolating and cable cover
PUR
Ʋ Colour
Black
Ʋ Fireprotectionclassification
UL94-V0
Connection cable, electrical data
Ʋ Wire cross-section
0.5 mm² (AWG 20)
Ʋ Wire resistance R
´
0.037Ω/m(0.012Ω/ft)
Ʋ Inductance L
i
´
0.6 µH/m (0.018 µH/ft)
Ʋ Capacitance Wire/Wire C
i
´
133 pF/m (40 pF/ft)
Ʋ Capacitance Wire/Screen C
i
´
215 pF/m (65 pF/ft)
Voltage supply
Operating voltage
Ʋ Non-Exinstrument
9 … 32 V DC
Ʋ Ex-iainstrument-Powersupply
FISCO model
9 … 17.5 V DC
Ʋ Ex-iainstrument-Powersupply
ENTITY model
9 … 24 V DC
Operating voltage with illuminated display and adjustment module
Ʋ Non-Exinstrument
13.5 … 32 V DC
Ʋ Ex-iainstrument-Powersupply
FISCO model
13.5 … 17.5 V DC
Ʋ Ex-iainstrument-Powersupply
ENTITY model
13.5 … 24 V DC
Powersupplyby/max.numberofsensors
Ʋ Fieldbus
max.32(max.10withEx)