Zeiss METROTOM Installation Instructions Manual - page 13
Installation Instructions
Page
13
METROTOM
2150
2205
3100
2223
600
min. 800
➊
➍
➎
➌
➀
➄
➃
➋
min. 700
min. 700
min. 700
850
600
➁
➅
min. 700
➂
➆
➂
➆
➀
➏
c
= The power connection and grounding terminal must be installed by the customer.
d
= The connecting point for the precleaned compressed air supply (6-10 bar) must be installed
by the customer.
e
= Teleservice
f
= Cable entry on control cabinet
g
= Compressed air supply on METROTOM (air hose entry from below, optionally from above)
h
= Computer station with 2x PC with monitor, keyboard and mouse
i
= Network connection
n
= Min. amount of free space required for maintenance. Recommended: 1350 mm
o
= The shaded area must be accessible for servicing so that bulky and heavy components can
be removed and installed if necessary. Transport to the left or right also must be possible.
p
= Cable: METROTOM - control cabinet approx. 5 m (cable entry from below)
q
= Cable: control cabinet - computer station approx. 13 m (cable entry from below)
r
= Cable: METROTOM - computer station approx. 20 m
s
= Cable: Network - control cabinet approx. 10 m (cable entry from below)
The computer station and control cabinet can also be installed in a separate room.
Note:
6. METROTOM installation dimensions and sample installation layout
3-wire outlet for
data station
(national voltage)
Power connection for
control cabinet
Net wt. of
METROTOM
Max. allowed
workpiece wt.
Control cabinet
approx.
ca. 8000 kg
50 kg
350 kg