2.6
Vibrations at the installation site
The permissible floor vibrations are illustrated in the following diagram(s).
Vibrations are often caused by heavy-duty machine tools, conveying equipment (crane or forklift), stamping and forging
machines located in adjacent rooms.
A measurement is required to assess the magnitude of the floor vibrations. If you would like to have this service performed
for you, contact Carl Zeiss in Oberkochen.
In any case, the measured result must be made available.
2.7
Floor loading capacity
We recommend that you consult a structural engineer for your statical calculations and any required floor reinforcements
(foundation etc.). The floor load exerted by the machine is specified in this brochure (Page 13).
10
1
0
10
20
30
40
50
100
1000
X axis, horizontal
Y axis, horizontal
Z axis, vertical
Frequency [Hz]
Acceleration [mm/ss]
Permissible Foundation Acceleration
Acceleration values above the curve require an additional
foundation