Fuse protection of line voltage
For 3/N/PE 400/230V line voltage:
– METROTOM with controller: Automatic cutout, 3-pole, C16A
– Data station: Automatic cutout, 1-pole C16A
For 3/PE 200 or 208V line voltage:
– METROTOM with controller: Automatic cutout, 3-pole, 15A
– Data station: Automatic cutout, 1-pole 15A
Power consumption
– METROTOM with controller: max. 3000VA
– Data station: max. 2000VA
3.4
EMC
EN 61326 Appendix A, Class A
Warning !
This is a class A device, and may therefore cause radio interference in the surrounding area. In such cases, the operator
may be required to take appropriate action and bear any costs thus incurred.
3.5
Installing the Connecting Cables
The cables can be protected by installing them in an electric raceway included in delivery. This raceway can accommodate
all cables and must be fastened to the floor. Or, as an alternative, the cables may also be installed in an underfloor cable
duct.
3.6
Stabilization of the power supply
If brief fluctuations occur in the root mean square value of the line voltage, a line voltage analysis must be performed.
Contact us if you require this service. We will be glad to advise you and recommend suitable measures. A separate stub
cable from the main distributing frame to the connecting point of METROTOM is sufficient in most cases.
3.7
Machine parameters
– Overvoltage category:
ΙΙΙ
– Pollution degree: 2
– Protection class: 1
3.8
Teleservice
You will require an Internet connection for Teleservice. For more details, please refer to the "Teleservice" brochure.
3.9
Network topology
The networking of the system components is performed by the ZEISS service engineer according to the valid specifications
during initial startup. These settings must not be modified by the customer. Any change could result in connection prob-
lems.
3.10 Computer operating system
If the required computers are provided by the customer, the operating system used must not show any limitations in com-
parison to the one used by ZEISS. Otherwise problems and additional costs may occur during startup.