3.2 Power supply on third-party tables
•
Connect all cables; if necessary, use a cable tie to relieve the tension on the pow-
er cable.
•
Fasten the PS-LED power supply with 4 screws.
•
Connect the multi-core plug connection between the head rest and the instrument
table.
•
Connect the main lamp cable to the upper part of the illumination unit.
WARNING!
•Care must be taken when mounting the guide rail and head rest to
ensure that there is no electrical connection to any metal part of the
table. Otherwise, it will be necessary to attach a ground conductor to
the head rest.
•The power supply must be connected with a ground conductor in any
case!
•Cables must not be trapped as this can cause a short circuit!
•Work on cables or device parts which are in connection with the
mainsmayonlybeconductedbyqualifiedpersonnel.
3.3 Instrument base with weight compensation facility
The weight of additional accessories mounted on the microscope can be compen-
sated using counterbalance springs. This keeps the height adjustment of the slit
lamp easy.
3.4 Setting the weight compensation facility
Turnthecontrollever(27)toitslowestpositionandthenslackenitslightlybya
quarter turn. Swivel the microscope and illumination to the side. Apply 1-3 springs
depending on the accessory.
3.5 Switching on the compensation facility
Turnanticlockwiseuntilthescrews(12)arecompletelyreleased.
3.6 Switching off the compensation facility
Turnthescrews(12)clockwiseuntilyoumeetresistance.Verifywhetherthemicro-
scope arm springs back downwards if you push it upwards with your hand. This will
only happen if the load is already at maximum. Generally, as many counterbalance
springs should be deactivated as necessary until this spring action occurs. The
weight compensation facility is set correctly once the illumination and microscope
with the mounted accessories weigh slightly more than the counterbalance springs.
3.7 Regulating the clearance of the slit width control
Thesmallscrewinthecenteroftherightcontrolknob(A)
allows you to regulate the friction of the turning movement
oftheseadjustingknobs.Turningitslightlytotheright(in)
makesitharder,turningitleft(out)makesiteasier.Itshould
at least be set so hard that the slit cannot close on its own.
A
3.8 Attaching/removing a camera
62. Lens barrel
63. Mirror and diaphragm component
64. Camera
65. Eyepiece
62
63
64
65
Remove the camera cover, attach the spacer ring to the camera and place the cam-
era on the lens barrel. If the camera is removed, the connection to the lens barrel
should be closed off with the cap intended for this purpose.
NOTE!
Consult your Haag-Streit representative for information on the support-
ed cameras and their connection rings.
4. Commissioning
ConnectthePS-LEDpowersupplyandFU01flashunittothemains.
4.1 Switching on the device
•
Throw the rocker switch on the PS-LED power supply and the main switch on the
FU01flashunit01I=’ON’.Thegreenlightsontherockerswitcheswilllightup
when the device is switched on.
•
Turn the rotating knob on the illumination control to a position between '1' and '10'.