Haag-Streit BX 900 Instructions For Use Manual - page 9
9
DEUTSCH
ENGLISH
FRANÇAIS
ITALIANO
ESPAÑOL
NEDERLANDS
PORTUGUÊS
SVENSKA
DEUTSCH
ENGLISH
FRANÇAIS
ITALIANO
ESPAÑOL
NEDERLANDS
PORTUGUÊS
SVENSKA
© HAAG-STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220449.04050 – 2015 – 06
2.10 Periphery or background illumination
58. Periphery/backgroundillumination(swiveling):Thisperiphery/
background illumination may only be used in conjunction with an
LI 01-plus LED illumination.
5 8
2.11 Power supply PS-LED
2.11.1 Models
59. Power supply PS-LED, for installation on third-party tables and units
60. Power supply PS-LED HSM 901 insert, for the HSM 901 instrument table
5 9
6 0
2.11.2 Description
61. Rocker switch
Switch in position 0 = ’OFF’, the power supply is discon-
nected from the mains, and the control lamp is unlit. Switch
in position I = ’ON’, the control lamp lights green.
6 1
O
I
O
a. Mainandfixationlampconnection
b. Connecting an illumination control from another
manufacturer(PowerJack2.1mm)
c. HAAG-STREIT connection for illumination control
IC01(USBminiB)
d. Typeplate(onhousing)
e. S1, pushbutton
f. L1,LEDdisplay(green/red)
g. Functionalearthconnection(M3)
h. 3-core mains connection
a b c d e f g h
3. Device assembly / installation
WARNING!
Thisdevicemustnotbemodifiedwithoutthemanufacturer'sapproval.
Installation and repairs may only be performed by trained specialists.
Contact your HAAG-STREIT representative for the installation, repair
andmodificationofthesystem.Youcanfindthecontactinformationat
www.haag-streit.com.
Theinitialinstallationofthedeviceisconductedbyanauthorizedservicetechni-
cian. The following adjustments can be conducted by the end user:
•
Regulating the clearance of the slit width control
•
Attaching/removingatonometer(seeseparateinstructionforuse)
•
Attaching/removing a camera
•
Adjusting the weight compensation
3.1 Power supply for instrument table HSM 901
NOTE!
HAAG-STREIT recommends connecting the aluminium tray to the
ground conductor connection on the power supply.
a. Connector orientation
b. Mainandfixationlampconnec-
tion
c. Type plate on housing
d. Mains connection
a
b
c
d
NOTE!
The 'Instrument table HSM 901' instruction for use must be followed.
01-IFU_BX900-7220449-04050_eng.indd 9