HAEFEL TEST ONYX User Manual - page 19
ONYX Electrostatic Discharge Simulator
15
5.4.2 Sub-menu 2
Open Setup
Open one of 8 stored setups. As factory default, 8 IEC tests are stored (L1 to L4
contact discharge and L1 to L4 air discharge).
This predefined tests can be overwritten by customer specific setups
Trigger Lock
"ON" locks trigger position, "OFF" disables trigger lock position.
When you select "ON" position, press the trigger once to begin a test sequence. If
pressed again the test will stop
Smart key
Program the smart key function according to your testing requirements: Alternate
Polarity, Reset Counter, Tip LED, Auxiliary trigger, Report, U-step
"ALTERNATE POL"
enables you to change the polarity via the smart key
"RESET COUNTER"
resets the counter to the pre-selected value
"Tip LED"
LED can be switched on/off by the smart key
"AUX. TRIGGER"
send a trigger signal to a PC via the optical link. Can be used for example for
triggering an USB-Camera which is connected to the PC
"REPORT"
If ONYX application software is used (running on PC which is connect to ONYX via
optical link), this function is used to jump automatically to the next test point on the
generated test report
"U-STEP"
Use the smart key to increase/decrease test levels: U-2kV, U-1kV, U-0.5kV,
U-0.1kV, U+0.1kV, U+0.5kV, U+1kV, U+2kV
Self test
Performs a self test on the ONYX simulator.
Please refer to section "Error Messages and Selftest"
Tip LED
Switch tip LED ON and OFF
Temp & RH
Displays environmental temperature, relative humidity and dew point in degrees
Fahrenheit or Celsius. Use up/down keys to toggle between degrees Fahrenheit and
Celsius
System Info
Displays firmware version, serial number, model, RC module values