HAEFEL TEST ONYX User Manual - page 26
22
ONYX Electrostatic Discharge Simulator
6.2 Selftest routine
Many ESD standards suggest determining whether the simulator is functioning correctly before testing takes place.
Running the Selftest routine (see section "Sub-menu 2"), all relevant parts of the unit are checked quick and easy.
6.2.1 Performing Selftest routine
Before performing the Selftest routine, ensure that the earth connection of the ESD Simulator is
connected to a good earth.
If Selftest routine is started without connection to earth, electrical shock may occur!
The test tip of the ONYX need to be short-circuited to earth during entire Selftest routine!
Contact ONYX tip to the ground reference plane of your test setup, then start Selftest from the menu
Failure to do this, some "FAILED" will result from the Selftest
The Selftest routine will check:
• Correct function of high voltage supply for both polarities
• Function of contact discharge relay
• Energy storage capacitor
• Discharge resistor
• Internal discharge circuit
A final report containing PASSED / FAILED information is displayed.
The Selftest duration time depends on the type of the RC module. For the default 150pF/330 Ohm RC module it
requires about 30s. For large capacitor values, it may take up to 3 minutes.
6.2.2 Troubleshooting if Selftest FAILED
In case that one or more FAILED occurred, please check the following:
• Ensure that the ONYX tip is short-circuited to ground during entire Selftest
• Fully charge the battery or insert a fully charged battery, perform Selftest again
• Switch Off the unit, remove the battery for 1 minute, after that perform Selftest again
• Perform Selftest using another test tip
• Perform Selftest using another RC module (if available)
If the unit is still not working properly, please contact our customer support.
For contact information, please see section "Corrections and additions".