HAEFEL TEST ONYX User Manual - 3 Technical Data
ONYX Electrostatic Discharge Simulator
3
3 Technical Data
3.1 Electrical data
3.1.1 ONYX basic version
ONYX 16 output voltage range
1kV to 16kV air discharge & contact discharge
ONYX 30 output voltage range
1kV to 30kV air discharge & contact discharge
Voltage resolution
100 V steps
Polarity
Positive / negative, automatic switching
Output voltage tolerance
± 5%
Holding time
> 5s
Default Discharge electrodes
Round tip for air discharge / sharp tip for contact discharge according
to IEC 61000-4-2 & ISO 10605
Default RC module
150pF / 330 Ohm according to IEC/EN 61000-4-2 edition 2,
waveform specification see table 1 below
RC module recognition
Automatic, values shown in display & test report
Controls
Touch screen, Trigger key, Smart key
Repetition frequency
Single, 0.1, 0.2, 0.5, 1, 2, 5, 10 and 20 Hz (see table 3)
Operating modes
Manual / continuous
Discharge detection / Threshold
OFF
(contact discharge mode only)
HV -10% … HV -90%
(contact discharge & air discharge single)
HV -20%
(fixed for air discharge auto)
Impulse counter
1 – 9999, pre-selectable
Temperature indicator
° Celsius or ° Fahrenheit
(selectable)
Humidity indicator
Relative humidity
Dew point indicator
° Celsius or ° Fahrenheit
(selectable)
Interface
Fibre-optic
Battery duration
Contact discharge 30kV, 1Hz, approx. 8h
Contact discharge 16kV, 1Hz, approx. 12h
Contact discharge 8kV, 1Hz, approx. 18h
Battery type
NiMH
Battery charger power requirements
100V – 240V, 47-63Hz, 17VA
Battery charging time
Approx. 3.5h
Level
Indicated
Voltage
First peak current
(± 15%)
Rise time tr
(± 25%)
Current @ 30ns
(± 30%)
Current @ 60ns
(± 30%)
[kV]
[A]
[ns]
[A]
[A]
1
2
7.5
0.8
4
2
2
4
15
0.8
8
4
3
6
22.5
0.8
12
6
4
8
30
0.8
16
8
X
16
60
0.8
32
16
X
* 30
112.5
0.8
60
30
Table 1 contact discharge current waveform parameters ONYX 16 / ONYX 30 according to
IEC 61000-4-2 edition 2
* ONYX 30 only