HAEFEL TEST ONYX User Manual - page 25
ONYX Electrostatic Discharge Simulator
21
6 Error Messages and Selftest
6.1 Error messages
ONYX monitors constantly all important parameters. Almost any major failure within one of the low
voltage or the high voltage circuits would be detected and reported immediately
When errors occur during operation of the ONYX, a warning message will appear.
If an error occurs, the ONYX will either request how to proceed, or following a warning, automatically switch off the unit
to avoid possible damage.
Fig. 15 error messages
Error Code Message
Possible cause
Action needed
0000 0001
Battery
voltage
error
Battery voltage too high
Battery voltage too low
Discharge battery via battery
charger and re-charge or
try other battery
Charge battery or try other battery
0000 0100
RC
missing
RC moudule missing or
damaged RC module
Switch off unit a place RC module
correctly or try another
RC module (if available)
0000 0200
TIP
missing
Tip missing or not placed correctly
Switch off unit and place tip
correctly
0000 0400
Charger
The charger is connected to the battery
while the ONYX is in operation
Switch off ONYX while charging
battery, or remove battery from
ONYX to charge
0000 4000
HV not
reached
This error may occur only in air
discharge mode, and indicates that
either a short circuit from tip to earth
has occurred, or that tip is too close to
the target
Check test setup
0000 8000
Timeout
HV low
The internal HV-capacitor discharge
circuit does not work correctly
Switch ONYX off and on again. If
error persists, unit need to be
repaired
0001 0000
Flash
memory
error
Error occurred while ONYX was writing
to memory. Setup will restore default
settings
Turn unit off using the power off
button on the display. Turn unit on
and try again
Table 4 error messages
If error codes appear without a displayed message, please take a note of the error code and contact our
service department. Contact details see section "Corrections and Additions"