I-SENS CareSens N Voice Manual - page 9
18 www.i-sens.com
www.caresens.co.nz 19
Step 9
In this mode all the test results stored in the meter can be
deleted. Please note that if you select YES, all the stored test
results will be deleted and can not be restored.
After the beeper mode is set,
press the S button to enter the
‘Test Result Reset’ mode. The
‘dEL’ character will blink on the
screen. Press or to alternate
between ‘YES’ or ‘no’. To delete
all the stored test results, press
the S button while the screen
displays ‘YES’. Then, all the test
results stored in the meter will
be deleted and the screen will be
similar to the picture on the right.
If you do not want to delete the results, press the S button while
the screen displays ‘no’. Then, the screen will return to step 2.
See page 14.
Setting the ‘Test Result Reset’
(Deleting all the saved test results)
Note:
At any stage, if the S button is pressed for 3 seconds, Time
and Date setting mode will finish and the meter will be turned
off. Press and hold to scroll through numbers quickly.
Checking the System
You may check your meter and test strips using the
CareSens Control Solution(control A and/or B).
The CareSens Control Solution contains a known amount
of glucose and is used to check that the meter and the
test strips are working properly.
The test strip vials have CareSens Control Solution ranges
printed on their labels. Compare the result displayed on
the meter to the CareSens Control Solution range printed
on the test strip vial.
Before using a new meter or a new vial of test strips, you may conduct
a control solution test following the procedure on pages 20-21.
Notes:
• UseonlytheCareSensControlSolution(availableforpurchase
separately).
• Checktheexpirationdatesprintedonthebottle.Whenyoufirst
open a control solution bottle, record the discard date (date
opened plus three (3) months) in the space provided on the label.
• Makesureyourmeter,teststrips,andcontrolsolutionareat
room temperature before testing. Control solution tests must be
done at room temperature (20-25˚C).
• Beforeusingthecontrolsolution,shakethebottle,discardthe
first few drops and wipe the tip clean.
• Closethecontrolsolutionbottletightlyandstoreata
temperature of 8-30˚C.
You may do a control solution test:
• Whenyouwanttopracticethetestprocedureusingthecontrol
solution instead of blood,
• Whenusingthemeterforthefirsttime,
• Wheneveryouopenanewvialofteststrips,
• Ifthemeterorteststripsdonotfunctionproperly,
• Ifyoursymptomsareinconsistentwiththebloodglucosetest