I-SENS CareSens N Voice User Manual - page 10
20 www.i-sens.com
www.i-sens.com 21
Note:
The CareSens Control Solution can be purchased
separately. Please contact your authorised i-SENS sales
representative.
Caution:
The range printed on the test strip vial is for the
CareSens Control Solution only. It has nothing to do with your
glucose level.
Step 3
The display segments will rotate clockwise on
the meter display and a test result will appear
after the meter counts down from 5 to 1.
After your control solution result appears on
the display, press
for3secondstill‘check’
appears on the display.
When‘check’isdisplayed,theresultisstoredin
themeter’smemorybutitisnotincludedinthe
averages.
Step 4
Compare the result displayed on the meter
to the range printed on the test strip vial.
Test result should fall within that range.
Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of this range.
Out of range results may occur in following situations:
If results continue to fall outside the range printed on the
test strip vial, the CareSens N Test Strip and Meter may not be
working properly. Do not use your system and contact your
authorised i-SENS sales representative.
Situations
•Whenthecontrolsolutionbottlewas
not shaken well,
•Whenthemeter,teststrip,orthecontrol
solution were exposed to high or low
temperatures,
•Whenthefirstdropofthecontrol
solution was not discarded or the tip of
the bottle was not wiped clean,
•Whenthemeterisnotfunctioning
properly.
•Whenthecontrolsolutionispastthe
expiration date printed on the bottle,
•Whenthecontrolsolutionispastits
discard date (the date the bottle was
opened plus three (3) months),
•Whenthecontrolsolutionis
contaminated.
Do This
Repeat the
control solution
test by referring
to the “Notes”
on page 18.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.