RAD B- 2.5 User manual - page 11
B-RAD 2.5 USER MANUAL
New World Technologies Inc.
Page • 9
V2014.07.03
Figure 2.4-1: Charging Status Display
2.4.1 Charging the RAD Li-Ion Battery Pack
Note: The temperature range for charging is 0˚C to 50˚C (32 ̊ F to 122 ̊ F).
To charge the RAD Battery:
1. Plug the RAD Battery Charger into the wall outlet.
Result: The Red Warning Light will turn on for one second and then the Green Status Light
will turn on for one second.
2. Align the RAD Battery with the RAD Battery Charger.
3. Slide the RAD Battery into place.
Result: The Green Status Light will flash while the RAD Battery is charging.
When the RAD Battery has been fully charged, the Green Status Light will stop flashing and
stay illuminated. Until the RAD Battery is removed from the RAD Charger, it will convert to
conservation mode which will maintain the battery charge at maximum capacity.
To remove the RAD Battery:
1. Slide the RAD Battery away from the RAD Charger.
2. Check that the RAD Battery is fully charged (refer to Section 2.3.2 – Checking RAD
Battery Charge).
2.4.2 Charging Errors
The Red Warning Light is on:
The RAD Battery is not charging because its temperature is not within the required
temperature range for charging. When the RAD Battery’s temperature moves within
the required range for charging, the Red Warning Light will turn off and charging will
commence.
The Red Warning Light is flashing:
The RAD Battery may be placed incorrectly on the RAD Battery Charger. Remove the
RAD Battery and replace it correctly on the RAD Battery Charger. If the Red Warning
Light continues to flash, the RAD Battery is defective; remove the RAD Battery
immediately.
If these problems continue, contact New World Technologies Inc. Technical Support (refer
to Section 5.0 – Contact Us) or your RAD Distributor.
Green Status
Light
Red Warning
Light