S.E. International Radiation alert Monitor 1000 EC Operation Manual - page 15
15
View
Preferences
Use View > Preferences menu to adjust the
settings in the Observer USB Preferences
(Figure 6(5)).
Chart Settings
Autosave and Clear Chart Every
Will automatically save and clear the chart at
the number of minutes selected. Auto Save
files are saved to the Observer USB Data
folder located in the Documents Library.
Use Audible Alarm
Will sound an alarm on the computer’s
sound system.
Grid Settings
Write to Grid Every
Will set how often a new data point will appear on the grid.
Autosave and Clear Grid Every
Will automatically save and clear the grid at the number of data points selected.
Additional Grid Data
Allows the setting of the units of measurement shown in the grid for
Value 1 and Value 2.
Cal Panel (Calibration Panel)
Located under View > Cal Panel, the Calibration Panel Window (Figure 6(6)) displays the
Serial Number and
Calibration Values. You can use the Calibration Panel to change
a number of settings on the instrument. The Monitor 1000
EC
should be in a mode screen or count screen when choosing
to
Update Settings, as some changes to the menu items via
the Cal Panel will not refresh if the instrument is displaying a
menu screen. Note: Update Settings must be chosen before any
changes made in the
Calibration Panel will be applied to the
Monitor 1000
EC
.
Calibration Information
Calibrations, such as an NIST Source Calibration, contain
information including the Serial Number, Calibration Date,
Deadtime, and Sensitivity (cpm per mR/hr).
CAUTION: Please note that changes to the Serial Number,
Calibration Date, Deadtime, and Sensitivity may void any
calibration of the instrument on record.
Figure 6(5)
Figure 6(6)