S.E. International Radiation alert Monitor 1000 EC Operation Manual - page 16
16
Alarm cpm and Alarm Enabled
To activate the alarm, enter the alarm threshold in the
Alarm cpm field, and check Alarm Enabled. To
disable the alarm, uncheck
Alarm Enabled. If the alarm has been activated and Alarm Enabled is then
unchecked, the alarm will continue to sound until
Update Settings has been selected.
Preset Counting Time (sec)
Enter the number of seconds to set the time for taking a timed count. The default is 600 seconds.
Use Data logging
Enables writing the collected data to the internal memory. Note: Set the time and date on the Monitor
1000
EC
before activating the data logging feature. The internal clock of the Monitor 1000
EC
can be set to
the time on a PC running the Observer USB Software by clicking Functions > Synchronize Monitor 1000
EC
to PC Clock. The Monitor 1000
EC
will beep twice when synchronization has completed unless
Quiet Mode is on.
Datalogging Interval (min)
Sets the interval for the data to be logged. The default is 1 minute and can only be changed through the software.
Circular Buffer
Allows the oldest collected data to be overwritten with the most recent data when the memory is full.
Clear Memory
Clears the data stored in the internal memory on the instrument.
Backlight On Time
Sets how many seconds the backlight will be on when pressing the backlight button.
Display Mode
Use to select from the available units of measurement
Contrast
Sets the contrast of the LCD display numbers and icons.
Auto-Averaging
This activates the auto-averaging feature. When unchecked, the Monitor 1000
EC
is in fast response mode.
Fast response mode is only recommended for use in high radiation fields.
Clicks Enabled
Enable or disable the clicks only.
Quiet Mode
Enable or disable all sounds on the Monitor 1000
EC
except the ALARM and OVER RANGE. You can still
enable clicks in
Quiet Mode.
Functions
Retrieve Memory
If the recording feature is activated and the internal clock is set, then the collected data can be retrieved from
the internal memory by selecting Functions > Retrieve Memory. Data is stored in a delimited text file, which
can be opened in your preferred spreadsheet software. The default data logging frequency is 1 minute. You can
change the value in the Calibration Panel window of the Observer USB Software.
Synchronize Monitor 1000
EC
to PC Clock
Sets the date and time on a PC running the Observer USB. The instrument will beep twice when
synchronization has completed.