Svantek SV 102 Quick Start Manual - page 10
SV 102 USER GUIDE
© SVANTEK
9
Press
+
to start the calibration measurement
When the calibration measurement is done, it shows
CAL. RESULT e.g. LEQ=114.22 dB
If you need to restart the calibration, press
+
again
Press
to accept the calibration result
Press
to confirm calibration factor and to store it
in TEDS memory of the microphone or, if you don’t want
to accept the calibration factor, press
and
push-
buttons to select NO; repeat the calibration
Instrument returns to the calibration window. Press
to return to the calibration menu, and take out
the
microphone
from
calibrator.
Calibrate
second
microphone in the similar way.
Now your SV 102 is ready to perform the measurements.
Press
+
push-buttons to see main view of display.
+