Svantek SV 102 Quick Start Manual - page 9
SV 102 USER GUIDE
© SVANTEK
8
Depending on the function being used, adjust other
necessary settings i.e. in:
DOSIMETER SETUP window (MENU > INPUT >
DOSIMETER SETUP)
SPECTRUM window (MENU > INPUT > SPECTRUM)
It is also possible to LOAD one of STANDARD SETUPS
(OSHA, MSHA, DOD, ACGIH, ISO 85, ISO 95) by selecting
proper option in FILE > LOAD SETUP > STANDARD SETUP
window and pressing
.
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
Go to MENU (
+
) > CALIBRATION > LEFT (RIGHT)
CHANNEL > BY MEASUREMENT
Enter the CAL. LEVEL according to the calibrator you use
(e.g. 113.9, 93.9)
Put the calibrated microphone into calibrator and switch
the calibrator on (SV 30A calibrator switches on
automatically)