Varian 947 Instruction Manual - Draf
Service Pendant Hand-Held Terminal
DRAF
T 2/4/02
11
[SHIFT]
INIT
EMIS
PUT
Sets a permanent new value for Emission Current. Enter XXXX
(the desired value), followed by the SHIFT key, then the
function key.
ZERO
TP TC
ATM
Initiates an atmosphere calibration of the Test Port TC.
[SHIFT]
ZERO
TP TC
ATM
Initiates a zero calibration of the Test Port TC.
ZERO
SY TC
ATM
Initiates an atmosphere calibration of the System TC.
[SHIFT]
ZERO
SY TC
ATM
Initiates a zero calibration of the System TC.
SET
TST TM
SHOW
Displays the value of the Auto Sequencer Test Time.
[SHIFT]
SET
TST TM
SHOW
Sets the value of the Auto Sequencer Test Time. Enter XXX
(the desired value), followed by the SHIFT key, then the
function key.
SET
RUF TM
SHOW
Displays the value of the Auto Sequencer Roughing Time.
[SHIFT]
SET
RUF TM
SHOW
Sets the value of the Auto Sequencer Roughing Time. Enter XXX
(the desired value), followed by the SHIFT key, then the
function key.
ON
A SEQ
OFF
Turns the Auto Sequencer function off.
[SHIFT]
ON
A SEQ
OFF
Turns the Auto Sequencer function on.
2
FIL SEL
1
Selects filament #1 in the Ion Source.
Table 2 Glossary of Input and Data Keys (continued)
CAUTION
System TC being set to atmosphere is not
a normal state for the leak detector. The
turbo pump must be turned off, and the
spectrometer tube must be vented. For
additional information, see the Model
947/948 or the Model 960 Operation
manuals.