Service Pendant Hand-Held Terminal
12
[SHIFT]
2
FIL SEL
1
Selects filament #2 in the Ion Source.
OUT V
SHOW
FIL #
Displays the active filament number.
[SHIFT]
OUT V
SHOW
FIL #
Displays the format (LOG or LIN) of the analog leak rate output
voltage.
[SHIFT]
RAW DTA
1
Displays a set of readings from the leak rate A to D conversion.
SET
REJ #
SHOW
Displays the active reject set point (1 through 4) and the value of
the reject set point.
[SHIFT]
SET
REJ #
SHOW
Sets the active reject set point. Enter a number from 1 through 4,
followed by the SHIFT key, then the function key.
SET
STD LK
SHOW
Displays the leak rate value of the internal calibrated leak. The leak
detector calibrates to this value.
[SHIFT]
SET
STD LK
SHOW
Sets the leak rate value to which the leak detector calibrates. Enter
X.XE
−
X or X.XE
−
XX (the desired value), followed by the SHIFT
key, then the function key.
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #1. Enter X.XE
−
X
(the desired value), then the function key.
[SHIFT]
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #3. Enter X.XE
−
X
(the desired value), followed by the SHIFT key, then the
function key.
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #2. Enter X.XE
−
X
(the desired value), followed by the SHIFT key, then the
function key.
[SHIFT]
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #4. Enter X.XE
−
X
(the desired value), followed by the SHIFT key, then the
function key.
Table 2 Glossary of Input and Data Keys (continued)