GL1000:
Replacing battery
The battery of the
GL1000
can be is exchanged after life cycle end.
Notes:
>
First read the installation instruction completely.
>
The case has to be opened to exchange the piggybacks.
>
This must be done very cautiously and carefully.
Caution:
When performing this operation be sure not to touch the top or bottom of
the boards (logger main board or piggybacks).
The battery is exchanged as follows:
1. First remove the two black decorative caps and the screws from the bottom cover
of the
GL1000
. The bottom cover contains the DSUB25 connector. Carefully pull
out the main circuit board.
2. You will find mounting location almost at the center of the main circuit board,
partly covered by the piggybacks (see green circle in Figure 2).
3. Remove the piggybacks for easier access to the battery.
4. Remove the battery carefully from the mounting location.
5. Insert the replacement battery. Look out for the correct polarity, + must be on top.
Please handle the contact spring with care. Do not bend it too much and make
sure the spring has contact to the new battery after replacement.
6. Place the
GL1000
main board back in the housing verifying that it is inserted
properly. This operation involves placing the housing on a table with its back side
(side with the bar code) facing up. Then the main board is inserted into the lower
guide rail with the piggybacks facing up.
7. It should be possible to slide the main board in the housing up to a few
millimeters from the end without forcing it in. Close the housing by applying light
pressure, and then secure it with the appropriate screw fasteners. The screws
should be secure but not excessively tight.
8. Please also attach the two black decorative caps.
Dispose of the removed battery according to the applicable laws (e.g. the Battery Law
in Germany).
2.2.9 Beep
Beep
The
GL1000
/
GL1010
has a speaker that acoustically alerts the user e.g. in case of a
trigger. Triggers and beep can be defined using the configuration program.
Because the opening in the housing has been eliminated, the signal tone of the
speaker is only audible to a limited extent in the
GL1010
.