Summary of VEGAFLEX 81

  • Page 1

    Quick setup guide tdr sensor for continuous level and interface measurement of liquids vegaflex 81 foundation fieldbus rod and cable probe document id: 47589.

  • Page 2: Contents

    2 contents vegaflex 81 • foundation fieldbus 47589-en-140324 contents 1 for your safety 1.1 authorised personnel ....................................................................................................... 3 1.2 appropriate use ................................................................

  • Page 3: 1  For Your Safety

    3 1 for your safety vegaflex 81 • foundation fieldbus 47589-en-140324 1 for your safety 1.1 authorised personnel all operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the devic...

  • Page 4

    4 1 for your safety vegaflex 81 • foundation fieldbus 47589-en-140324 electromagnetic compatibility instrumentsinfour-wireorex-d-iaversionaredesignedforuseinan industrial environment. Nevertheless, electromagnetic interference from electrical conductors and radiated emissions must be taken into acco...

  • Page 5: 2  Product Description

    5 2 product description vegaflex 81 • foundation fieldbus 47589-en-140324 2 product description 2.1 configuration thetypelabelcontainsthemostimportantdataforidentificationand use of the instrument: 2 1 13 14 15 16 12 11 5 3 6 4 7 8 9 10 fig. 1: layout of the type label (example) 1 instrument type 2 ...

  • Page 6

    6 2 product description vegaflex 81 • foundation fieldbus 47589-en-140324 • scanthedatamatrixcodeonthetypelabeloftheinstrumentor • enter the serial number manually in the app.

  • Page 7: 3 Mounting

    7 3 mounting vegaflex 81 • foundation fieldbus 47589-en-140324 3 mounting 3.1 general instructions for use of the instrument protect your instrument against moisture penetration through the fol- lowing measures: • use the recommended cable (see chapter "connectingtopower supply") • tighten the cable...

  • Page 8

    8 3 mounting vegaflex 81 • foundation fieldbus 47589-en-140324 plastic vessel/glass vessel the guided microwave principle requires a metallic surface on the processfitting.Therefore,inplasticvessels,etc.,useaninstru- mentversionwithflange(fromdn50)orplaceametalsheet (ø>200mm/8in)beneaththeprocessfit...

  • Page 9

    9 3 mounting vegaflex 81 • foundation fieldbus 47589-en-140324 h d d h dn40 ... Dn150 > dn150 ... Dn200 150 100 _ _ fig. 4: mounting socket whenweldingthesocket,makesurethatthesocketisflushwiththe vessel top. 1 2 fig.5:socketmustbeinstalledflush 1 unfavourable installation 2 socketflush-optimuminsta...

  • Page 10

    10 4 connecting to power supply vegaflex 81 • foundation fieldbus 47589-en-140324 4 connecting to power supply 4.1 connecting the voltage supply and signal output are connected via the spring- loaded terminals in the housing. Connection to the display and adjustment module or to the interface adapte...

  • Page 11

    11 4 connecting to power supply vegaflex 81 • foundation fieldbus 47589-en-140324 fig.7:connectionsteps5and6-doublechamberhousing 6. Insert the wire ends into the terminals according to the wiring plan information: solidcoresaswellasflexiblecoreswithwireendsleevesareinsert- eddirectlyintotheterminal...

  • Page 12

    12 4 connecting to power supply vegaflex 81 • foundation fieldbus 47589-en-140324 1 2 ( ) (-) 1 5 0 1 0 1 + 6 7 8 bus 2 3 4 5 fig.8:electronicsandterminalcompartment,singlechamberhousing 1 voltage supply, signal output 2 contactpinsforthedisplayandadjustmentmoduleorinterfaceadapter 3 simulation swit...

  • Page 13: Module

    13 5 set up with the display and adjustment module vegaflex 81 • foundation fieldbus 47589-en-140324 5 set up with the display and adjustment module 5.1 insert display and adjustment module the display and adjustment module can be inserted into the sensor andremovedagainatanytime.Youcanchooseanyoneo...

  • Page 14

    14 5 set up with the display and adjustment module vegaflex 81 • foundation fieldbus 47589-en-140324 1 2 fig.11:insertionofthedisplayandadjustmentmoduleintothedoublechamber housing 1 in the electronics compartment 2 intheconnectioncompartment(withex-d-iaversionnotpossible) note: ifyouintendtoretrofi...

  • Page 15

    15 5 set up with the display and adjustment module vegaflex 81 • foundation fieldbus 47589-en-140324 max. Adjustment inthismenuitem,youcanenterthemax.Adjustmentforthelevel. Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. The distance refers to t...

  • Page 16

    16 5 set up with the display and adjustment module vegaflex 81 • foundation fieldbus 47589-en-140324 linearization alinearizationisnecessaryforallvesselsinwhichthevesselvolume doesnotincreaselinearlywiththelevel-e.G.Ahorizontalcylindri- cal or spherical tank, when the indication or output of the vol...

  • Page 17: 6 Supplement

    17 6 supplement vegaflex 81 • foundation fieldbus 47589-en-140324 6 supplement 6.1 technical data electromechanical data - version ip 66/ip 67 and ip 66/ip 68; 0.2 bar cable gland m20x1.5or½npt wire cross-section (spring-loaded terminals) Ʋ massive wire, cord 0.2 … 2.5 mm² (awg 24 … 14) Ʋ stranded w...

  • Page 18

    18 notes vegaflex 81 • foundation fieldbus 47589-en-140324.

  • Page 19

    19 notes vegaflex 81 • foundation fieldbus 47589-en-140324.

  • Page 20

    Printing date: vega grieshaber kg am hohenstein 113 77761 schiltach germany 47589-en-140324 all statements concerning scope of delivery, application, practical use and operat- ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subjec...