Image optimization
Choosing a lookup table
• Select Palette in the Image window of the
scanned image (Fig. 19).
The Color Palette window appears.
• In the Color Palette List panel, click on the
Range Indicator item (Fig. 20).
The scanned image appears in a false-color
presentation (Fig. 19).
If the image is too bright, it appears red on the
screen. Red = saturation (maximum).
If the image is not bright enough, it appears blue
on the screen. Blue = zero (minimum).
Adjusting the laser intensity
• Set the Pinhole to 1 Airy Unit (Fig. 21).
• Set the Detector Gain high.
• When the image is saturated, reduce AOTF
transmission in the Excitation panel (Fig. 21)
using the slider to reduce the intensity of the
laser light at the specimen.
Adjusting gain and offset
• Increase the Amplifier Offset until all blue
pixels disappear, and then make it slightly
positive (Fig. 21).
• Reduce the Detector Gain until the red pixels
only just disappear.
Fig. 19
Image window
Fig. 20
Color Palette window
Fig. 21
Scan Control window, Channel settings