Zeiss LSM 510 Brief Operating Manual - page 9
07/05
9
• Select the appropriate filters and activate the
channels.
• Click the Excitation button to select the laser
lines and set the attenuation values
(transmission in %) in the displayed window.
For the configuration of the beam path, please
refer to the application-specific configurations
depending on the used dyes and markers and the
existing instrument configuration.
• Clicking on the Spectr button
opens the Detection Spectra &
Laser Lines … window (Fig. 13) to
display the activated laser lines for
excitation (colored vertical lines) and
channels (colored horizontal bars).
• Clicking on the Config button
opens the Track Configurations
window (Fig. 14) to load, store or
delete track configurations.
• For storing a new track configuration enter a
desired name in the first line of the
Configurations list box and click an Store.
• For loading an existing configuration select it in
the list box and click on Apply.
• For deleting an existing configuration select it in
the list box and click on Delete.
Setting for multi track configuration in
Channel Mode
The Multi Track function permit several tracks to
be defined as one configuration (Channel Mode
Configuration) for the scan procedure, to be
stored under any name, reloaded or deleted.
The maximum of four tracks with up to 8 channels
can be defined simultaneously and then scanned
one after the other. Each track is a separate unit
and can be configured independently of the other
tracks with regard to channels, Acousto-Optical
Tunable Filters (AOTF), emission filters and dichroic
beam splitters.
• Select Channel Mode if necessary (Fig. 15).
Fig. 13
Detection Spectra & Laser Lines …
window
Fig. 14
Track Configurations window
Fig. 15
Configuration Control window for
Multi Track