Zeiss LSM 510 Brief Operating Manual - Setting The Microscope
6
07/05
• Click on the Laser button to open the Laser Control window.
• Select the appropriate Laser Unit by clicking on
the name of it.
• Click on the Standby button to switch required
laser(s) to Standby.
• When status is Ready click on On button.
• Set Output [%] so that the Tube Current is
about 4 A (50% of the output power).
Setting the microscope
Changing between direct observation or laser scanning
The VIS, TV and LSM buttons switch the beam path and indicate which beam path has been set in the
binocular tube of the microscope:
• Click on the VIS button to set the microscope for direct observation via the eyepieces
of the binocular tube, lasers are off.
• Click on the TV button to set the microscope camera observation (if connected) via
camera adapter of the binocular tube.
• Click on the LSM button to set the microscope screen observation via laser excitation
using the LSM 510 and software evaluation.
Setting the microscope and storing the settings
• Click on the VIS button for direct observation.
• Click on the Micro button in the Acquire subordinate toolbar to open the
Microscope Control window of the used microscope.
The Microscope Control window appears (Fig. 10).
Fig. 9
Laser Control window