Accutome ACCUPEN User Manual

Other manuals for ACCUPEN: User Manual
Manual is about: HANDHELD TONOMETER

Summary of ACCUPEN

  • Page 1

    Accupen user’sguide accu pen ® handheld tonometer 24-3002 rev-i.

  • Page 2

    Accutome accupen user’s guide ii www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe federal law restricts this device to sale by or on the order of a physician. Federal communications commission (fcc) unintentional emitter per fcc part 15 this device...

  • Page 3

    Iii (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com chapter 1: introduction 1 accupen overview 1 features 1 measurements 2 about this manual 2 chapter 2: safety 3 safety information 3 safety issues to consider when using the accupen 3 indications ...

  • Page 4

    Accutome accupen user’s guide iv www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe storage 14 troubleshooting 14 chapter 5: specifications 16 overview 16 physical specifications 16 environmental specifications 17 measurement accuracy 17 emc complian...

  • Page 5

    V (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com table of contents list of figures figure 1.1 - accupen tonometer 1 figure 2.1 - clean with canned air 6 figure 3.1 - accupen unpacked 8 figure 3.3 - action control buttons and lcd 9 figure 3.2 - ba...

  • Page 6

    1 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com chapter 1: introduction 1 introduction accupen overview the accutome accupen pictured below has all the features that make it easy to obtain extreme accuracy and improved patient outcomes. Figure 1...

  • Page 7

    Accutome accupen user’s guide 2 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe measurements the high accuracy of the accupen measurements is provided by the following: • high-resolution, real-time waveform analysis • high-speed signal digitalizat...

  • Page 8

    3 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com chapter 2: safety 2 safety safety information the section lists: • specific safety precautions associated with the accupen • general safety precautions safety issues to consider when using the accu...

  • Page 9

    Accutome accupen user’s guide 4 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe disposal in europe the european community (ec) has issued two directives; 91/157/eec and 93/86/eec. These directives are implemented by each member country in a differ...

  • Page 10

    5 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com safety precautions there are several areas in the use of the accupen that require special attention, as they may pose a safety threat. The accupen has an enclosure rated degree of protection of ip3...

  • Page 11

    Accutome accupen user’s guide 6 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe to clean the accupen, perform the following steps: 1. Remove accutip cover from the tonometer, if one is installed. 2. Submerge the tip of the accupen in isopropyl alc...

  • Page 12

    7 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com the device covers must not be removed except by qualified personnel. There are no user controls inside the unit. To avoid injury, do not operate the accupen without protective covers. The system is...

  • Page 13

    Accutome accupen user’s guide 8 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe chapter 3: getting started 3 getting started overview the accupen is designed to be used in multiple medical settings and can be rested on a surface, such as a counter...

  • Page 14

    9 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com caution: use only the style and type of battery specified. Using another style or type of battery may cause damage to the product and invalidate the warranty. Battery installation caution: the batt...

  • Page 15

    Accutome accupen user’s guide 10 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe calibration the accupen is calibrated during the manufacturing process, so there is no need to calibrate the unit prior to using it for the first time. Recalibration ...

  • Page 16

    11 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com getting started figure 3.6 - calibration probe down position 5. Wait for a beep, then a calibration number will be displayed on the lcd. 6. Press the action control button again to exit calibratio...

  • Page 17

    Accutome accupen user’s guide 12 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe how to take a measurement 1. Touch and hold the up and down control buttons on the accupen simultaneously for 2 to 3 seconds to reset all measurements, averages, and ...

  • Page 18

    13 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com getting started table 3.1 below provides the iop correction values. Table 3.1 - iop correction values corneal thickness (micrometers) correction values (mmhg) 405 7 425 6 445 5 465 4 485 3 505 2 5...

  • Page 19

    Accutome accupen user’s guide 14 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe general maintenance maintenance that should be performed on the accupen consists of activities such as keeping surfaces free of dust and dirt and storing in a dry and...

  • Page 20

    15 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com table 4.1 - continued accupen troubleshooting information symptom probable cause correction multiple variable readings battery is low replace battery (see chapter 3) mechanical or electronic damag...

  • Page 21

    Accutome accupen user’s guide 16 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe chapter 5: specifications 5 specifications overview this section provides the physical and operational specifications of the accupen. Physical specifications table 5....

  • Page 22

    17 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com environmental specifications table 5.2 below lists the accupen system operating and storage values for temperature and humidity. Table 5.2 - environmental specifications temperature operating +10 ...

  • Page 23

    Accutome accupen user’s guide 18 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe table 5.4 - guidance and manufacturer’s declaration - electromagnetic emissions the accupen is intended for use in the electromagnetic environment specified below. Th...

  • Page 24

    19 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com specifications table 5.6 - guidance and manufacturer’s declaration - electromagnetic immunity the accupen is intended for use in the electromagnetic environment specified below. The customer or th...

  • Page 25

    Accutome accupen user’s guide 20 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe table 5.7 - recommended separation distances between portable and mobile rf communications equipment and the accupen the accupen is intended for use in an electromagn...

  • Page 26

    21 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com chapter 6: warranty & repairs 6 warranty & repairs warranty accutome, inc., warrants its new equipment to be free from defects in workmanship or materials. Any product that is proven to be defecti...

  • Page 27

    Accutome accupen user’s guide 22 www.Accutome.Com • (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe nonreturnable merchandise accutome, inc., will not authorize a return for: • merchandise held longer than 60 days replacement parts table 6.1 below lists items that a...

  • Page 28

    23 (800) 979-2020 us & ca • (610) 889-0200 international • +31 (0) 485-350300 europe • www.Accutome.Com warranty & repairs.

  • Page 29

    Us and canada (800) 979-2020 international (610) 889-0200 info@accutome.Com www.Accutome.Com.