Accutome ACCUPEN User Manual - page 23
Accutome AccuPen User’s Guide
18
www.accutome.com
• (800) 979-2020 US & CA • (610) 889-0200 International • +31 (0) 485-350300 Europe
Table 5.4 - Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The AccuPen is intended for use in the electromagnetic environment specified below. The
customer or the user of the AccuPen should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment
Guidance
RF Emissions CISPR 11* Group 1
The AccuPen uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and not likely
to cause any interference in nearby
electronic equipment
RF Emissions CISPR 11* Class B
The AccuPen is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
N/A
Table 5.5 - Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The AccuPen is intended for use in the electromagnetic environment specified below. The
customer or the user of the AccuPen should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
Discharge
(ESD)
+/- 6kV
Contact
+/- 6kV
Contact
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
IEC 61000-4-2
+/- 8kV Air
+/- 8kV Air
Power
frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.