Accutome ACCUPEN User Manual - page 24
19
(800) 979-2020 US & CA • (610) 889-0200 International • +31 (0) 485-350300 Europe •
www.accutome.com
Specifications
Table 5.6 - Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The AccuPen is intended for use in the electromagnetic environment specified below. The
customer or the user of the AccuPen should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Conducted
RF IEC
61000-4-6
3 Vrms 150
kHz to 80
MHz
N/A
Portable and mobile RF
communications equipment should
be used no closer to any part of the
AccuPen, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 1.2√P 800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
metres (m). Field strengths from fixed
RF transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference
may occur in the vicinity of
equipment marked with the following
symbol:
Radiated
RF IEC
61000-4-3
3 V/m
80MHz to
2,5 GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the AccuPen is used
exceeds the applicable RF compliance level above, the AccuPen should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the AccuPen.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.