Bacharach Microprocessor Module Replacement Kit Instruction Manual - page 10
Microprocessor Module Replacement Kit for Multi-Zone Leak Monitors
10
3015-5547 Rev 4
8. Reload Setup Parameters
Reload the previously saved setup parameters and configuration data by following the steps below.
a. With the monitor in either Warm Up Mode (flashing green LED) or Sampling Mode (solid green LED), access
the Setup Menu by pressing the ENTER key twice.
Figure 14. Sampling Mode Screen
b. Scroll to the “SYSTEM” option and press ENTER.
Figure 15. Main Setup Screen
c. Use the UP and DOWN arrow keys to select the LOCATION setting and press ENTER. (Note: This is the
location of the Multi-Zone unit, and NOT the location of the individual zones.) Use the LEFT and RIGHT
arrow keys to select the letter to be changed, and use the UP and DOWN arrow keys to choose the new letter
or number for that position. Press ENTER to save the changes, or ESC to abort the changes.
Figure 16. HGM Setup Screens